DFMS HK monitoring {| TABLE !RSDB !!PDS Keyword !!HK-Mnemonic old !!Description !!HK Type !!Std./Ext. Pack. !!Monitoring if !!Sampling frequency !!Transmission frequency S/C !!Range !!Criti-cality !!Action, if outside range !!Used for regulation !!Remarks ! |- | || || || || || || || || || || || || || |- | || || || || || || || || || || || || || |- |NRNAD101 ||ROSINA_DFMS_ISHEATER_FLAG ||DFMS IS Heater ||Ion source heater Enabled ||Flag ||Std. ||No measurement with IS heater operation ||2s ||1 min ||On/Off ||3 ||9 || || |- |NRNAD101 ||ROSINA_DFMS_GCU1_ENABLED_FLAG ||DFMS Flag GCU 1 ||GCU 1 heater Enabled ||Flag ||Std. ||Measurement with GCU operation ||200 ms /synchronized with measurement ||1 min ||On/Off, switching ||3 ||10 || || |- |NRNAD101 ||ROSINA_DFMS_GCU2_ENABLED_FLAG ||DFMS Flag GCU 2 ||GCU 2 heater Enabled ||Flag ||Std. ||Measurement with GCU operation ||200 ms /synchronized with measurement ||1 min ||On/Off, switching ||3 ||10 || || |- |NRNAD101 ||ROSINA_DFMS_COV_MOT_DIS_FLAG ||DFMS Flag Cover ||Cover motor Disabled ||Flag ||Std. ||Measurement ||2s / synchronized with measurement ||1 min ||On ||3 ||TBD || || |- |NRNAD101 ||ROSINA_DFMS_NOP_FLAG ||DFMS Flag NOP ||Non operational mode Enabled ||Flag ||Std. ||Measurement ||2s / synchronized with measurement ||1 min ||On ||TBD || || || |- |NRNAD101 ||ROSINA_DFMS_CMD_ERROR_FLAG ||DFMS Flag Cmd Err ||Command Error ||Flag ||Std. ||Measurement ||2s / synchronized with measurement ||1 min || ||TBD || || || |- |NRNAD101 ||ROSINA_DFMS_HV_SAFE_FLAG ||DFMS FLag HV Safe ||HV Safe Status ||Flag ||Std. ||Measurement ||2s / synchronized with measurement ||1 min ||Off ||3 ||TBD || || |- |NRNAD101 ||ROSINA_DFMS_HV_REDUCED_FLAG ||DFMS Flag HV Reduced ||HV Reduced Status ||Flag ||Std. ||Measurement ||2s / synchronized with measurement ||1 min ||Off ||3 || || || |- |NRNAD103 ||ROSINA_DFMS_HK_COUNTER ||DFMS HK counter ||number of HK packets sent || ||Std. ||DPU generated || ||1 min ||not applicable ||n.a. || || || |- |NRNAD104 ||ROSINA_DFMS_CMD_COUNTER ||DFMS Cmd counter ||number of DFMS commands processed || ||Std. ||DPU generated || ||1 min ||not applicable ||n.a. || || || |- |NRNAD105 ||ROSINA_DFMS_CMD_ERROR_CNT ||DFMS Cmd Error cnt ||number of wrong DFMS commands || ||Std. ||DPU generated || ||1 min ||not applicable ||n.a. || || || |- |NRNAD106 ||ROSINA_DFMS_CMD_ERROR_POS ||DFMS Cmd Error pos || || ||Std. ||DPU generated || ||1 min ||not applicable ||n.a. || || || |- |NRNAD107 ||ROSINA_DFMS_SCI_COUNTER ||DFMS Science counter ||number of DFMS science packets sent by DPU || ||Std. ||DPU generated || ||1 min ||not applicable ||n.a. || || || |- |NRNAD108 ||ROSINA_DFMS_SCI_ERROR_COUNTER ||DFMS Science Error counter || || ||Std. ||DPU generated || ||1 min ||not applicable ||n.a. || || || |- |NRNAD109 ||ROSINA_DFMS_SCI_ERROR_POSITION ||DFMS Science Error position || || ||Std. ||DPU generated || ||1 min ||not applicable ||n.a. || || || |- |NRNAD10A ||ROSINA_DFMS_MODE_TABLE ||DFMS S/W mode || || ||Std. ||DPU generated || ||1 min ||not applicable ||n.a. || || || |- |NRNAD10B ||ROSINA_DFMS_STATUS ||DFMS S/W status || || ||Std. ||DPU generated || ||1 min ||not applicable ||n.a. || || || |- |NRNAD10C ||ROSINA_DFMS_LEDA_CEM_ACCU_TIME ||Leda/CEM Accu Time ||Accumulation time of MCP or CEM detector || ||Std. ||DPU generated || ||1 min ||not applicable ||n.a. || || || |- |NRNAD10D ||ROSINA_DFMS_ACCU_SCAN_COUNTS ||Accu/Scan Counts ||Total number of counts on detector || ||Std. ||DPU generated || ||1 min ||not applicable ||n.a. || || || |- |NRNAD10E ||ROSINA_DFMS_MODE_TABLE ||MEP CEM DTA || || ||Std. ||DPU generated || ||1 min ||not applicable ||n.a. || || || |- |NRNAD10F ||ROSINA_DFMS_SEQUENCE_TABLE ||Spare 1 || || ||Std. ||DPU generated || ||1 min ||not applicable ||n.a. || || || |- |NRNAD110 ||ROSINA_DFMS_MASS_HK ||Spare 2 || || ||Std. ||DPU generated || ||1 min ||not applicable ||n.a. || || || |- |NRNAD111 ||ROSINA_DFMS_SENT_CMDS_COUNTER ||Spare 3 || || ||Std. ||DPU generated || ||1 min ||not applicable ||n.a. || || || |- |NRNAD112 ||ROSINA_DFMS_MEP_MAIN_CURRENT ||MEP I MEP ||MEP main current monitor (all loads except ASP, FDP) ||Measured ||Std. ||Measurement ||2s / synchronized with measurement ||1 min ||< 0.4 A ||2 ||TBD || ||Monitored for power consumption |- |NRNAD113 ||ROSINA_DFMS_MEP_FLOAT_CURRENT ||MEP I Float ||MEP floating voltage current monitor (ASP, FDP) ||Measured ||Std. ||Measurement ||2s / synchronized with measurement ||1 min ||< 0.3 A ||2 ||TBD || ||Monitored for power consumption |- |NRNAD114 ||ROSINA_DFMS_MEP_MESHED_GRID ||MEP MG ||Meshed grid voltage ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD115 ||ROSINA_DFMS_MEP_ION_SOURCE_BIAS ||MEP ISB ||Ion source box voltage ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD116 ||ROSINA_DFMS_MEP_GCU1_INPRESSURE ||MEP GCU PIn 1 ||Gas calibration unit 1 input pressure (high pressure gauge) ||Measured ||Std. ||Measurement with GCU operation ||200 ms /synchronized with measurement ||1 min ||TBD ||3 ||6 || ||Monitored for extensive gas loss |- |NRNAD117 ||ROSINA_DFMS_MEP_GCU1_CURRENT ||MEP GCU I 1 ||Gas calibration unit 1 valve heater current ||Measured ||Std. ||Measurement with GCU operation ||200 ms /synchronized with measurement ||1 min || ||3 ||5 || || |- |NRNAD118 ||ROSINA_DFMS_MEP_GCU1_OUTPRESSURE ||MEP GCU POut 1 ||Gas calibration unit 1 output pressure (Mini Pirani) ||Measured ||Std. ||Measurement with GCU operation ||200 ms /synchronized with measurement ||1 min ||TBD range ||3 ||6-Jul ||GCU operation ||Monitored for uncontrolled pressure rise |- |NRNAD119 ||ROSINA_DFMS_MEP_GCU2_INPRESSURE ||MEP GCU PIn 2 ||Gas calibration unit 2 input pressure (high pressure gauge) ||Measured ||Std. ||Measurement with GCU operation ||200 ms /synchronized with measurement ||1 min ||TBD ||3 ||6 || ||Monitored for extensive gas loss |- |NRNAD11A ||ROSINA_DFMS_MEP_GCU2_CURRENT ||MEP GCU I 2 ||Gas calibration unit 2 valve heater current ||Measured ||Std. ||Measurement with GCU operation ||200 ms /synchronized with measurement ||1 min || ||3 ||5 || || |- |NRNAD11B ||ROSINA_DFMS_MEP_GCU2_OUTPRESSURE ||MEP GCU POut 2 ||Gas calibration unit 2 output pressure (Mini Pirani) ||Measured ||Std. ||Measurement with GCU operation ||200 ms /synchronized with measurement ||1 min ||TBD range ||3 ||6-Jul ||GCU operation ||Monitored for uncontrolled pressure rise |- |NRNAD11C ||ROSINA_DFMS_MEP_GCU1_TEMP1 ||MEP GCU Temp 11 ||Gas calibration unit 1 temperature sensor 1 ("tank side") ||Measured ||Std. ||Measurement with GCU operation ||200 ms /synchronized with measurement ||1 min ||-30 to +60 DegC (TBC) ||3 ||5 || || |- |NRNAD11D ||ROSINA_DFMS_MEP_GCU1_TEMP2 ||MEP GCU Temp 12 ||Gas calibration unit 1 temperature sensor 2 ("sensor side") ||Measured ||Std. ||Measurement with GCU operation ||200 ms /synchronized with measurement ||1 min ||-30 to +60 DegC (TBC) ||3 ||5 || || |- |NRNAD11E ||ROSINA_DFMS_MEP_GCU2_TEMP1 ||MEP GCU Temp 21 ||Gas calibration unit 2 temperature sensor 1 ("tank side") ||Measured ||Std. ||Measurement with GCU operation ||200 ms /synchronized with measurement ||1 min ||-30 to +60 DegC (TBC) ||3 ||5 || || |- |NRNAD11F ||ROSINA_DFMS_MEP_GCU2_TEMP2 ||MEP GCU Temp 22 ||Gas calibration unit 2 temperature sensor 2 ("sensor side") ||Measured ||Std. ||Measurement with GCU operation ||200 ms /synchronized with measurement ||1 min ||-30 to +60 DegC (TBC) ||3 ||5 || || |- |NRNAD120 ||ROSINA_DFMS_MEP_CEM_CURRENT ||MEP CEM I ||Strip current flowing in the channeltron ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||TBD range ||3 ||TBD || || |- |NRNAD121 ||ROSINA_DFMS_MEP_CEM_ANALOG_HG ||MEP CEM ANA ||Current collected at channeltron anode ||Measured ||Std. ||Not applicable ||science data ||science data ||not applicable ||n.a. ||no action ||CEM gain control ||will be high gain analog output for FS |- |NRNAD122 ||ROSINA_DFMS_MEP_COV_MOT_CURR ||MEP MOT I ||Cover motor current ||Measured ||Std. ||No measurement with cover operation ||2s ||1 min || ||2 ||4 || || |- |NRNAD123 ||ROSINA_DFMS_MEP_HALL_COUNTER ||Hall Counter ||Cover Hall Counter ||Flag ? ||Std. ||No measurement with cover operation ||2s ||1 min || ||2 ||TBD || || |- |NRNAD124 ||ROSINA_DFMS_MEP_MOT_COUNTER ||Motor Counter ||Cover Motor Counter ||Flag ? ||Std. ||No measurement with cover operation ||2s ||1 min || ||2 ||TBD || || |- |NRNAD125 ||ROSINA_DFMS_MEPCOVER_OPEN_FLAG ||MEC_MOT_STAT COV OPEN ||Cover Open Status ||Flag ||Std. ||No measurement with cover operation ||2s ||1 min || ||2 ||TBD || || |- |NRNAD125 ||ROSINA_DFMS_MEP_COVER_CLOSE_FLAG ||MEC_MOT_STAT COV CLOSED ||Cover Closed Status ||Flag ||Std. ||Measurement without cover operation ||2s / synchronized with measurement ||1 min ||True/False ||2 ||TBD || || |- |NRNAD126 ||ROSINA_DFMS_CEM_THRESHOLD ||MEP DAC THR ||CEM threshold setting ||DAC ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||2/set flag low or high|| ||transmitted only as off/low/high/ok flag |- |NRNAD127 ||ROSINA_DFMS_CEM_FRONT ||MEP HVCEM DAC ||CEM front side high voltage setting ||DAC ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- 50V (TBC) ||2 ||1 /2 /set flag low or high || ||transmitted only as off/low/high/ok flag |- | || || || || || || || || || || || || || |- |NRNAD128 ||ROSINA_DFMS_MEP_CEM_CURRENT ||MEP HVCEM I ||CEM high voltage current of regulation loop ||Measured ||Std. ||Measurement ||2s / synchronized with measurement ||1 min ||Nominal value +/-TBD % or TBD range ||3 (TBC) ||TBD || || |- |NRNAD129 ||ROSINA_DFMS_ISC_HV ||MEP HVACC DAC ||Acceleration voltage setting ||DAC ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||1 / 2 /set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD12A ||ROSINA_DFMS_ISC_HV_CURRENT ||MEP HVACC I ||Acceleration voltage current of regulation loop ||Measured ||Std. ||Measurement ||2s / synchronized with measurement ||1 min ||Nominal value +/-TBD % or TBD range||3 (TBC) ||TBD || || |- |NRNAD12B ||ROSINA_DFMS_IONSRC_TEMPRATURE ||MEP IS Temp ||Temperature measured for the ion source ||Measured ||Std. ||Measurement ||2s ||1 min ||-80 to +500 DegC ||2 ||set flag low or high ||IS heater operation || |- |NRNAD12C ||ROSINA_DFMS_MEP_ACCEL_VOLT_ERR ||MEP HVACC ERR ||Acceleration voltage error output of regulation loop ||Measured ||Std. ||Measurement ||2s / synchronized with measurement ||1 min ||Nominal value +/-TBD % or TBD range||3 (TBC) ||TBD || || |- |NRNAD12D ||ROSINA_DFMS_CEM_FRONT_ERROR ||MEP HVCEM ERR ||CEM high voltage error output of regulation loop ||Measured ||Std. ||Measurement ||2s / synchronized with measurement ||1 min ||Nominal value +/-TBD % or TBD range||3 (TBC) ||TBD || || |- |NRNAD12E ||ROSINA_DFMS_MEP_CEM_GND ||MEP CEM GND ||Reference Ground of channeltron output ||Measured ||Std. ||Measurement ||2s / synchronized with measurement ||2 min ||n/a || || || || |- |NRNAD12F ||ROSINA_DFMS_FIL1_BIAS ||ISC FIL 1 Bias ||Filament 1 biasing ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||3 / set flag low or high || || |- |NRNAD130 ||ROSINA_DFMS_FIL2_BIAS ||ISC FIL 2 Bias ||Filament 2 biasing ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||3 / set flag low or high || || |- |NRNAD130 ||ROSINA_DFMS_ISC_HV_FLAG ||Fil 2 Bias ISC HVx ||HV Ion source Enabled ||Flag ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD130 ||ROSINA_DFMS_ISC_TL_FLAG ||Fil 2 Bias TLx ||HV Transfer Lens Enabled ||Flag ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD131 ||ROSINA_DFMS_FIL1_EMISSION ||ISC FIL 1 E ||Filament 1 emission current ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||3 ||3 / set flag low or high || || |- |NRNAD131 ||ROSINA_DFMS_ISC_SL_FLAG ||Fil 1 Emis SLx ||HV Source Lens Enabled ||Flag ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD131 ||ROSINA_DFMS_ISC_SE_FLAG ||Fil 1 Emis Sex ||HV Source Entrance Slit Enabled ||Flag ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD132 ||ROSINA_DFMS_FIL_EMISSION ||ISC FIL 1 I ||Filament 1 current ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||3 ||TBD / set flag low or high || || |- |NRNAD132 ||ROSINA_DFMS_FIL1_ENABLED_FLAG ||Fil 1 Curr Fil1 ||Filament 1 Enabled ||Flag ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD132 ||ROSINA_DFMS_FIL2_ENABLED_FLAG ||Fil 1 Curr Fil2 ||Filament 2 Enabled ||Flag ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD133 ||ROSINA_DFMS_FIL2_EMISSION ||ISC FIL 2 E ||Filament 2 emission current ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/-TBD % ||3 ||TBD /set flag low or high || || |- |NRNAD134 ||ROSINA_DFMS_ISC_FIL2_CURRENT ||ISC FIL 2 I ||Filament 2 current ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/-TBD % ||3 ||TBD / set flag low or high || || |- |NRNAD135 ||ROSINA_DFMS_ELLECTRON_REPELLER ||ISC ERP ||Electron repeller voltage ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||set flag low or high || || |- |NRNAD136 ||ROSINA_DFMS_ION_REPELLER1 ||ISC IRP1 ||Ion repeller 1 voltage ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD137 ||ROSINA_DFMS_ION_REPELLER2 ||ISC IRP2 ||Ion repeller 2 voltage ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD138 ||ROSINA_DFMS_ION_SUPRESSOR ||ISC ISP ||Ion suppressor voltage ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD139 ||ROSINA_DFMS_SOURCE_LENS_LEFT ||ISC SLL VMON ||Source lens left output voltage ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||3 ||TBD /set flag low or high || || |- |NRNAD13A ||ROSINA_DFMS_SOURCE_LENS_RIGHT ||ISC SLR VMON ||Source lens right output voltage ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/-TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD13B ||ROSINA_DFMS_SOURCE_EXIT_SLIT ||ISC SES VMON ||Source exit slit output voltage ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD13C ||ROSINA_DFMS_BETA_DEFLECTOR ||ISC SEB VMON ||Beta deflector output voltage ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD13D ||ROSINA_DFMS_TRANSFER_LENS_RIGHT ||ISC TLR VMON ||Transfer lens right output voltage ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD13E ||ROSINA_DFMS_TRANSFER_LENS_LEFT ||ISC TLL VMON ||Transfer lens left output voltage ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/-TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD13F ||ROSINA_DFMS_ASP_HV_BIAS ||ASP E HVB DAC ||Internal ASP HV Bias setting ||DAC ||Std. ||Measurement ||2s / synchronized with measurement ||1 min || ||4 ||no action || || |- |NRNAD140 ||ROSINA_DFMS_ASP_HV_BIAS_CURR ||ASP E HVB I ||Internal ASP HV Bias current of regulation loop ||Measured ||Std. ||Measurement ||2s / synchronized with measurement ||1 min ||TBD (provided by T. Magoncelli) ||3 ||TBD || || |- |NRNAD141 ||ROSINA_DFMS_POSTACCEL_VOLTAGE ||ASP E FDP DAC ||HV FDP reference setting ||DAC ||Std. ||Measurement ||2s / synchronized with measurement ||1 min || ||4 ||no action || || |- |NRNAD142 ||ROSINA_DFMS_ASPD_REF_VOLT ||ASP D VREF ||ASP D reference voltage ||Measured ||Std. ||Measurement ||2s / synchronized with measurement ||1 min || ||4 ||no action || || |- |NRNAD143 ||ROSINA_DFMS_ENTR_SLIT_SWITCH2 ||ASP D ESS2 DAC ||Entrance slit switch condenser 1 setting ||DAC ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD144 ||ROSINA_DFMS_ENTR_SLIT_SWITCH1 ||ASP D ESS1 DAC ||Entrance slit switch condenser 2 setting ||DAC ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD145 ||ROSINA_DFMS_ZOOM_QUAD2 ||ASP D Z2Q DAC ||Zoom quadruple 2 setting ||DAC ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD146 ||ROSINA_DFMS_ELSTAT_INNER_PLATE ||ASP C ESAIT DAC ||Electrostatic analyzer inner toroid fine setting ||DAC ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value+/-TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD146 || ||ASP C ESAIT DAC ||Electrostatic analyzer ??? ||DAC ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD147 ||ROSINA_DFMS_ELSTAT_OUTER_PLATE ||ASP C ESAOT DAC ||Electrostatic analyzer outer toroid fine setting ||DAC ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value+/-TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD148 ||ROSINA_DFMS_ELSTAT_INNER_TOROID ||ASP C ESAC DAC ||Electrostatic analyzer coarse setting ||DAC ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD149 ||ROSINA_DFMS_ZOOM_QUAD1 ||ASP C Z1QV DAC ||Zoom quadruple 1 setting ||DAC ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD14A ||ROSINA_DFMS_HEXAPOLE ||ASP A HP DAC ||Hexapole setting ||DAC ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD14B ||ROSINA_DFMS_RATION_QUAD ||ASP A RQ DAC ||Rotation quadruple setting ||DAC ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD14C ||ROSINA_DFMS_MATSUDA_PLATE ||ASP A MP DAC ||Matsuda plates setting ||DAC ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- TBD % ||2 ||set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD14D ||ROSINA_DFMS_CEM_REPELLER ||ASP A CEM DAC ||CEM repeller setting ||DAC ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- 10V (TBC) ||2 ||2/set flag low or high || ||transmitted only as off/low/high/ok flag |- |NRNAD150 ||ROSINA_DFMS_MAGNET_TEMPERATURE ||ASP B MAG Temp ||Temperature measured on the sector magnet ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data || ||4 || ||mass settings (calculation of VACCL) ||transmitted as value |- |NRNAD152 ||ROSINA_DFMS_MCP_BACKVOLTAGE1 ||FDP MCP Back 1 ||MCP back side high-voltage 1 ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- 10V (TBC) ||2 ||1/2/set flag low or high || ||Nominal value for Back >= -300V and <= 40V ! (TBC) |- |NRNAD153 ||ROSINA_DFMS_MCP_BACKVOLTAGE2 ||FDP MCP Back 2 ||MCP back side high-voltage 2 ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- 10V (TBC) ||2 ||1/2/set flag low or high || ||Nominal value for Back >= -300V and <= 40V ! (TBC) |- |NRNAD154 ||ROSINA_DFMS_RDP_TEMPERATURE ||FDP RDP Temp ||Temperature measured in detector proximity electronics ||Measured ||Std. ||Measurement ||2s / synchronized with measurement ||1 min ||> +50 DegC ||3 ||Deactivate RDP heater ||RDP heater operation ||If LEDA temperature measurement failes, LEDA temperature will be replaced by RDP temperature |- |NRNAD155 ||ROSINA_DFMS_LEDA_TEMPERATURE ||FDP LEDA Temp ||Temperature measured internally in LEDA512 chip behind MCP ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min ||< +20-TBD DegC ||3 ||Activate RDP heater if RDP temperature < +50DegC ||RDP heater operation || |- |NRNAD156 ||ROSINA_DFMS_FARADAY_SIGNAL ||FDP Cup Out ||Output signal of faraday cup ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||n/a ||4 || || || |- |NRNAD157 ||ROSINA_DFMS_FARADAY_LOW_FLAG ||FDP Low Range Status ||Status of Faraday cup electrometer low range relay ||Flag ||Std. ||Measurement ||2s / synchronized with measurement ||1 min ||-3 to +3 V ||4 || || || |- |NRNAD158 ||ROSINA_DFMS_FARADAY_MED_FLAG ||FDP Medium Range Status ||Status of Faraday cup electrometer medium range relay ||Flag ||Std. ||Measurement ||2s / synchronized with measurement ||1 min ||-3 to +3 V ||4 || || || |- |NRNAD159 ||ROSINA_DFMS_MCP_FRONTVOLTAGE ||FDP MCP HV ||MCP front side high-voltage ||Measured ||Std. ||Measurement ||2s /synchronized with measurement ||1 min/with science data ||Nominal value +/- 50V (TBC) ||2 ||1/2/set flag low or high || || |- |NRNAD15A ||ROSINA_DFMS_FDPC_CURRENT ||FDP FDPC_I ||MCP high voltage convertor primary current ||Measured ||Std. ||Measurement ||2s / synchronized with measurement ||1 min ||< TBD upper limit (provided by T. Magoncelli) ||2 ||1 || || |- |NRNAD15B || ||FDP Cup Rep V ||Repeller voltage on Faraday cup ||Measured ||Std. ||Measurement ||2s / synchronized with measurement ||2 min ||TBD ||4 || || || |- |NRNAD15C ||ROSINA_DFMS_FAR_REP_ENABL_FLAG ||FDP_STAT FC Rep. Ena ||Faraday Cup Repeller Enabled ||Flag ||Std. ||Measurement ||2s /synchronized with measurement ||with science data ||On/Off ||2 ||set flag low or high || ||transmitted only as off/on flag |- |NRNAD15C ||ROSINA_DFMS_RDP_HEATER_FLAG ||FDP_STAT RDP Heater ||Remote detector pack heater Enabled ||Flag ||Std. ||Measurement ||2s / synchronized with measurement ||1 min ||On/Off ||3 ||8 || ||RDP heating successful ? |- |NRNAD15C ||ROSINA_DFMS_MCPHV_ACC_BSY ||FDP_STAT Accu Busy ||MCP/LEDA Accumulation busy ||Flag ||ext. ||Measurement ||2s / synchronized with measurement ||5 min ||On/Off ||4 || || || |- |NRNAD15C ||ROSINA_DFMS_MCPHV_ENABLED_FLAG ||FDP_STAT MCP HV Ena ||MCP HV Enabled ||Flag ||ext. ||Measurement ||1 min / synchronized with measurement ||5 min ||On/Off ||4 || || || |- |NRNAD15C ||ROSINA_DFMS_MCPHV_REDUCED_FLAG ||FDP_STAT MCP HV Red ||MCP HV Reduced Status ||Flag ||ext. ||Measurement ||1 min / synchronized with measurement ||5 min ||On/Off ||4 || || || |- |NRNAD15C ||ROSINA_DFMS_MCPHV_SAFE_FLAG ||FDP_STAT HV Safe ||MCP HV Safe Status ||Flag ||ext. ||Measurement ||1 min / synchronized with measurement ||5 min ||On/Off ||4 || || || |- |NRNAD161 ||ROSINA_DFMS_MEPA_TEMPERATURE ||MEP A Temp ||Temperature measured in the MEP-A board (command and data handler) ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min ||-30 to +60 DegC (TBC) ||3 ||TBD || || |- |NRNAD162 ||ROSINA_DFMS_MEPN_TEMPERATURE ||MEP N Temp ||Temperature measured in the MEP-N board (low voltage power supply) ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min ||-30 to +60 DegC (TBC) ||3 ||TBD || || |- |NRNAD163 ||ROSINA_DFMS_COV_MOT_TEMPERATURE ||MEP MOT Temp ||Cover motor temperature ||Measured ||Ext. ||No measurement with cover operation ||2s ||5 min || ||2 ||4 || || |- |NRNAD164 ||ROSINA_DFMS_MEPD_REFVOLT ||MEP D REF RED ||MEP D reference reduced ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD165 ||ROSINA_DFMS_MEPE_REFVOLT ||MEP E REF RED ||MEP E reference reduced ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD166 ||ROSINA_DFMS_CEM_ANALOG_LG ||MEP DAC PWA ||Spare || || || || || || || || || ||will be low gain analog output of CEM for FS |- |NRNAD167 ||ROSINA_DFMS_MEP_DAC_STIMAMP ||MEP DAC STIM AMP ||Spare || || || || || || || || || || |- |NRNAD168 ||ROSINA_DFMS_MEP_REFVOLT ||MEP VREF ||MEP reference voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD169 ||ROSINA_DFMS_MEPG_TEMPERATURE ||MEP G Temp ||Temperature measured in the MEP-G board (Ion source control) ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min ||-30 to +60 DegC (TBC) ||3 ||TBD || || |- |NRNAD16A ||ROSINA_DFMS_ISC_SLL_REFVOLT ||ISC SLL RMON ||Source lens left reference program voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD16B ||ROSINA_DFMS_ISC_SLR_REFVOLT ||ISC SLR RMON ||Source lens right reference program voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD16C ||ROSINA_DFMS_ISC_SES_REFVOLT ||ISC SES RMON ||Source exit slit reference program voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD16D ||ROSINA_DFMS_ISC_SEB_REFVOLT ||ISC SEB RMON ||Beta deflector reference program voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD16E ||ROSINA_DFMS_ISC_TLR_REFVOLT ||ISC TLR RMON ||Transfer lens right reference program voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD16F ||ROSINA_DFMS_ISC_TLL_REFVOLT ||ISC TLL RMON ||Transfer lens left reference program voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD170 ||ROSINA_DFMS_MEP_H_TEMPERATURE ||MEP H Temp ||Temperature measured in the MEP-H board (transfer optics voltage reference) ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min ||-30 to +60 DegC (TBC) ||3 ||TBD || || |- |NRNAD171 ||ROSINA_DFMS_ISC_HV_REFVOLT ||ISC HV RMON ||Raw -2.8kV reference voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD172 ||ROSINA_DFMS_ISC_SEB_DRIVEVOLT ||ISC SEB DMON ||Beta deflector opto-coupler drive voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD173 ||ROSINA_DFMS_ISC_SES_DRIVEVOLT ||ISC SES DMON ||Source exit slit opto-coupler drive voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD174 ||ROSINA_DFMS_ISC_SLR_DRIVEVOLT ||ISC SLR DMON ||Source lens right opto-coupler drive voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD175 ||ROSINA_DFMS_ISC_SLL_DRIVEVOLT ||ISC SLL DMON ||Source lens left opto-coupler drive voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD176 ||ROSINA_DFMS_ISC_TLR_DRIVEVOLT ||ISC TLR DMON ||Transfer lens right opto-coupler drive voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD177 ||ROSINA_DFMS_ISC_TLL_DRIVEVOLT ||ISC TLL DMON ||Transfer lens left opto-coupler drive voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD178 ||ROSINA_DFMS_ISC_HV_DRIVEVOLT ||ISC HV DMON ||Raw -2.8kV converter drive voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD179 ||ROSINA_DFMS_ISC_PWR_DRIVEVOLT ||ISC PWR DMON ||Analog power monitor ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD17A ||ROSINA_DFMS_ISC_HV ||ISC HV VMON ||Raw -2.8kV voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD17B ||ROSINA_DFMS_ISC_HV_CURRENT ||ISC HV I ||Raw -2.8kV converter current ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min ||TBD (provided by T. Magoncelli) ||3 ||TBD || || |- |NRNAD17C ||ROSINA_DFMS_ESA_COARSE_DAC ||ASP C_ESAC_DAC ||analyzer part, electrostatic analyzer coarse voltage, DAC value ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||6 min || || || || || |- |NRNAD17D ||ROSINA_DFMS_ASP_HVBIAS_REFVOLT ||ASP E HVB ERR ||Internal ASP HV Bias monitor ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD17E ||ROSINA_DFMS_ASP_HVFDP_ERROR ||ASP E FDP ERR ||HV FDP reference error output of regulation loop ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD17F ||ROSINA_DFMS_ASP_Z2QV_ERROR ||ASP D Z2QV ERR ||Zoom quadruple 2 vertical error output of regulation loop ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD180 ||ROSINA_DFMS_ASP_Z2QH_ERROR ||ASP D Z2QH ERR ||Zoom quadruple 2 horizontal error output of regulation loop ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD181 ||ROSINA_DFMS_ASP_ESS2_ERROR ||ASP D ESS2 ERR ||Entrance slit switch condenser 2 error output of regulation loop ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD182 ||ROSINA_DFMS_ASP_ESS1_ERROR ||ASP D ESS1 ERR ||Entrance slit switch condenser 1 error output of regulation loop ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD183 ||ROSINA_DFMS_ASP_ESSAO_ERROR ||ASP C ESAO ERR ||Electrostatic analyzer error output of regulation loop ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD184 ||ROSINA_DFMS_ASP_ESAI_ERROR ||ASP C ESAI ERR ||Electrostatic analyzer error output of regulation loop ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD185 ||ROSINA_DFMS_ASP_Z1QV_ERROR ||ASP C Z1QV ERR ||Zoom quadruple 1 vertical error output of regulation loop ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD186 ||ROSINA_DFMS_ASP_Z1QH_ERROR ||ASP C Z1QH ERR ||Zoom quadruple 1 horizontal error output of regulation loop ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD187 ||ROSINA_DFMS_ASP_CEMREP_ERROR ||ASP A CEM ERR ||CEM Repeller error output of regulation loop ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD188 ||ROSINA_DFMS_ASP_HP_ERROR ||ASP A HP ERR ||Hexapole error output of regulation loop ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD189 ||ROSINA_DFMS_ASP_RQ_ERROR ||ASP A RQ ERR ||Rotation quadruple error output of regulation loop ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD18A ||ROSINA_DFMS_ASP_MP_ERROR ||ASP A MP ERR ||Matsuda plates error output of regulation loop ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD18B ||ROSINA_DFMS_ASPC_TEMPERATURE ||ASP B ESA Temp ||Temperature measured in the ASP-C board (power supply electrostatic analyzer) ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min ||-30 to +60 DegC (TBC) ||3 ||TBD || || |- |NRNAD18C ||ROSINA_DFMS_ASPB_TEMPERATURE ||ASP B ASP Temp ||Temperature measured in the ASP-B board (digital control) ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min ||-30 to +60 DegC (TBC) ||3 ||TBD || || |- |NRNAD18D ||ROSINA_DFMS_ASP_15V_MONITOR ||ASP B 15V Mon ||ASP 15 V supply voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD18E ||ROSINA_DFMS_ASP_HV_ENABL_FLAG ||ASP_STAT HV On ||Internal ASP HV Bias turned on ||Flag ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD18E ||ROSINA_DFMS_ASP_HV_RED_FLAG ||ASP_STAT ASP A ||ASP-A HV output Enabled ||Flag ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD18E ||ROSINA_DFMS_ASPA_HV_ENABLED_FLAG ||ASP_STAT ASP C ||ASP-C HV output Enabled ||Flag ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD18E ||ROSINA_DFMS_ASPC_HV_ENABLED_FLAG ||ASP_STAT ASP D ||ASP-D HV output Enabled ||Flag ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD18E ||ROSINA_DFMS_ASPD_HV_ENABLED_FLAG ||ASP_STAT ASP E ||ASP-E HV output Enabled ||Flag ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD18E ||ROSINA_DFMS_ASPE_HV_ENABLED_FLAG ||ASP_STAT HV Safe ||ASP HV Safe Status (inhibits ASP HV Ena and ASP A Ena through ASP E) ||Flag ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD18E ||ROSINA_DFMS_ASP_HV_SAFE_FLAG ||ASP_STAT HV Red ||ASP HV Reduced Status Enabled (HV reduced by 8) ||Flag ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- |NRNAD18F ||ROSINA_DFMS_FDP_GROUND_VOLTAGE ||FDP Ground ||Floating detector pack 0 Volt refererence ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || 0V +/- 0.5V (TBC) ||3 ||1 / 2 || || |- |NRNAD190 ||ROSINA_DFMS_FDP_5NEG_VOLTAGE ||FDP -5V ||Floating detector pack -5 V supply voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || 0V +/- 0.5V (TBC) ||3 ||1 / 2 || || |- |NRNAD191 ||ROSINA_DFMS_FDP_5POS_VOLTAGE ||FDP +5V ||Floating detector pack +5 V supply voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || 0V +/- 0.5V (TBC) ||3 ||1 / 2 || || |- |NRNAD192 ||ROSINA_DFMS_FDP_75POS_VOLTAGE ||FDP +7.5V ||Floating detector pack +7.5 V supply voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min ||+7,5V +/- 0.5V (TBC) ||3 ||1 / 2 || || |- |NRNAD193 ||ROSINA_DFMS_FDP_75NEG_VOLTAGE ||FDP -7.5V ||Floating detector pack -7.5 V supply voltage ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min ||+7,5V +/- 0.5V (TBC) ||3 ||1 / 2 || || |- |NRNAD194 ||ROSINA_DFMS_FDPA_TEMPERATURE ||REF43 Temp ||Temperature measured in reference circuit on FDP-A ||Measured ||Ext. ||Measurement ||2s / synchronized with measurement ||5 min || ||4 ||no action || || |- | || || || || || ||Measurement ||synchronized with measurement ||1 min ||-3 to +3 V ||4 ||TBD || || |} General remark: In case of filament 2 operation, the monitoring procedure for FIL 1 Emis and FIL 2 Emis have to be exchanged ! {| TABLE !Legend !! !! ! |- |Criticality ||Description ||Montioring, action items |- |1 ||emergency case ||immediate switch off, emergency handling by operator |- |2 ||short time effects ||immediate and appropriate action by DPU |- |3 ||long time effects ||observation for a certain time, appropriate action by DPU |- |4 ||not critical at all ||no monitoring, no actions |} {| TABLE !Action !!Description ! |- |1 ||Turn DFMS off, turn DFMS on again, return to current sensor state. If problem still exists, turn DFMS completely off until appropriate reaction by operator. |- |2 ||Turn DFMS off, turn DFMS on again, repeat current measurement mode. If problem still exists, turn DFMS completely off until appropriate reaction by operator. |- |3 ||Turn off filament, try it again. If problem still exists, turn DFMS completely off until appropriate reaction by operator. |- |4 ||Stop cover operation, TBD further action. |- |5 ||Stop GCU operation, TBD further action. |- |6 ||Turn DFMS off, TBD further action. |- |7 ||Stop GCU operation, if problems still exists, turn DFMS completely off, TBD further action. |- |8 ||Stop RDP heater operation, try it again. If problems still exists, TBD further action. |- |9 ||Stop IS heater operation, try it again. If problems still exists, TBD further action. |- |10 ||Stop GCU operation, try it again. If problems still exists, TBD further action. |} RTOF HK monitoring {| TABLE !RSDB !!HK-Mnemonic old !!PDS keyword !!Description !!HK Type !!Std./Ext. Pack. !!Monitoring if !!Sampling frequency !!Transmission frequency S/C !!Range !!Criti-cality !!Action, if outside range !!Used for regulation !!Remarks ! |- |0 ||NRNAR101 || ||RTOF Std. HK Header || ||std || ||2s ||1 min || || || || || |- |1 || || || || ||std || ||2s ||1 min || || || || || |- | || || ||Spare || ||std || ||2s ||1 min || || || || || |- | || || || ||Flag ||std || ||2s ||1 min || || || || || |- | || || || ||Flag ||std || ||2s ||1 min || || || || || |- |2 ||NRNAR103 ||ROSINA_RTOF_SCI_HK_COUNT ||RTOF HK counter ||Value ||std || ||2s ||1 min || || || || || |- |4 ||NRNAR104 ||ROSINA_RTOF_SCI_CMD_COUNT ||RTOF Cmd counter ||Value ||std || ||2s ||1 min || || || || || |- |6 ||NRNAR105 ||ROSINA_RTOF__SCI_CMD_ERR_CNT ||RTOF Cmd Error counter ||Value ||std || ||2s ||1 min || || || || || |- |8 ||NRNAR106 ||ROSINA_RTOF__SCI_CMD_ERR_POS ||RTOF Cmd Error position ||Value ||std || ||2s ||1 min || || || || || |- |10 ||NRNAR107 ||ROSINA_RTOF_SCI_COUNT ||RTOF Science counter ||Value ||std || ||2s ||1 min || || || || || |- |12 ||NRNAR108 ||ROSINA_RTOF_SCI_ERR_COUNT ||RTOF Science Error counter ||Value ||std || ||2s ||1 min || || || || || |- |14 ||NRNAR109 ||ROSINA_RTOF_SCI_ERR_POSITION ||RTOF Science Error position ||Value ||std || ||2s ||1 min || || || || || |- |16 ||NRNAR10A ||ROSINA_RTOF_SCI_MODE ||RTOF S/W mode ||Value ||std || ||2s ||1 min || || || || || |- |18 ||NRNAR10B ||ROSINA_RTOF_SCI_STAT ||RTOF S/W status ||Value ||std || ||2s ||1 min || || || || || |- |20 ||NRNAR10C ||ROSINA_RTOF_SCI_MODE_NUMBER ||RTOF Science Mode number ||Value ||std || ||2s ||1 min || || || || || |- |22 ||NRNAR10D || ||Status Bits 1 || ||std || ||2s ||1 min || || || || || |- | || ||ROSINA_RTOF_SCI_PSU_9KV_FLAG ||power supply 9 kV enable status ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_PSU_70V_FLAG ||power supply 70 V enable status ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_PSU_OS_MCP_FLAG ||power supply ortho MCP enable status ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_PSU_SS_MCP_FLAG ||power supply storage MCP enable status ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_PSU_HM_POW_FLAG ||power supply hard mirror enable status ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_PSU_PULS_FLAG ||power supply pulser enable status ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || || ||Spare ||Flag ||std || ||2s ||1 min || || || || || |- | || ||ROSINA_RTOF_SCI_ETS_RAM_T_ACT_F ||ETS Ram Test Active ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETS_RAM_T_STA_F ||ETS Ram Test Status ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETS_RAM_T_DATA ||ETS Ram Test Data ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETS_INPUT_STAT_F ||Input Selection: Ion, Gas or Calibrator (Input Select) ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||N.U ||ETS Output Enable ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETS_CAL_TRIG_F ||Equivalent Time Sampling_Controller ||Flag ||std || ||2s ||1 min || ||4 || || || |- |24 ||NRNAR10E ||ROSINA_RTOF_SCI_LOWERREADADDRESS ||ETS Lower Readout Address ||Value ||std || ||2s ||1 min || ||4 || || || |- |26 ||NRNAR10F ||ROSINA_RTOF_SCI_UPPERREADADDRESS ||ETS Upper Readout Address ||Value ||std || ||2s ||1 min || ||4 || || || |- |28 ||NRNAR110 || ||Status Bits 2 || ||std || ||2s ||1 min || ||4 || || || |- | || || ||Spare ||Flag ||std || ||2s ||1 min || || || || || |- | || ||ROSINA_RTOF_SCI_FEC_FIL2_SS_FLAG ||Filament emission controller Fil 2 Storage Status ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_FEC_FIL1_SS_FLAG ||Filament emission controller Fil 1 Storage Status ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_FEC_FIL2_OS_FLAG ||Filament emission controller Fil 2 Ortho Status ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_FEC_FIL1_OS_FLAG ||Filament emission controller Fil 1 Ortho Status ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_FECI_STATUS_FLAG ||Filament emission controller I Status ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_FEC_EH_OS_FLAG ||Filament emission controller EH Ortho Status ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_FEC_EH_SS_FLAG ||Filament emission controller EH Storage Status ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_LRA_BIT16 ||Equivalent Time Sampling_Postprocessor (ETS Lower Readout Address Bit 16) ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_URA_BIT16 ||Equivalent Time Sampling_Postprocessor (ETS upper Readout Address Bit 16) ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETSLOWERRANGE_F ||Equivalent Time Sampling_Postprocessor (ETS Lower Range) ||Flag ||std || ||2s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETSUPPERRANGE_F ||Equivalent Time Sampling_Postprocessor (ETS Upper Range) ||Flag ||std || ||2s ||1 min || ||4 || || || |- |30 ||NRNAR111 || ||Spare || ||std || ||2s ||1 min || || || || || |- |32 ||NRNAR112 ||ROSINA_RTOF_SCI_OS_FIHEAT_I ||Filament emission controller_Orthosource_Filament Heater_current [mA] (MC_FEC_ION_FIHEAT_I) ||measured ||std ||FEC is running ||2s ||1 min ||? ||2 ||11 || ||nominal range TBD ??? |- |34 ||NRNAR113 ||ROSINA_RTOF_SCI_SS_FIHEAT_I ||Filament emission controller_Storagesource_Filament Heater_current [mA] (MC_FEC_GAS_FIHEAT_I ) ||measured ||std ||FEC is running ||2s ||1 min ||? ||2 ||11 || ||nominal range TBD ??? |- |36 ||NRNAR114 ||ROSINA_RTOF_SCI_OS_REP_V_A ||Orthosource repeller voltage #A ||measured ||std || ||2s ||1 min ||? ||3 || || || |- |38 ||NRNAR115 ||ROSINA_RTOF_SCI_OS_REP_V_B ||Ortho source repeller voltage #B ||measured ||std || ||2s ||1 min ||? ||3 || || || |- |40 ||NRNAR116 ||ROSINA_RTOF_SCI_SS_REP_V_A ||Storage source repeller voltage #A ||measured ||std || ||2s ||1 min ||? ||3 || || || |- |42 ||NRNAR117 ||ROSINA_RTOF_SCI_SS_REP_V_B ||Storage source repeller voltage #B ||measured ||std || ||2s ||1 min ||? ||3 || || || |- |44 ||NRNAR118 ||ROSINA_RTOF_SCI_OS_FIL_V ||Orthosource filament voltage ||measured ||std || ||2s ||1 min ||? ||3 || || || |- |46 ||NRNAR119 ||ROSINA_RTOF_SCI_SS_FIL_V ||Storage source filament voltage ||measured ||std || ||2s ||1 min ||? ||3 || || || |- |48 ||NRNAR11A ||ROSINA_RTOF_SCI_SS_FIL_I ||Filament emission controller_Storagesource_Filament emission_current [uA] (MC_Filament emission controller_Storage_FIL_I) ||measured ||std || ||2s ||1 min ||? +-15uA ||3 ||27 || || |- |50 ||NRNAR11B ||ROSINA_RTOF_SCI_OS_ENT1_V ||Filament emission controller_Orthosource_Entrance1_voltage (MC_Filament emission controller_Ortho_ENT1_V) ||measured ||std || ||2s ||1 min ||? +-5V ||3 ||28 || || |- |52 ||NRNAR11C ||ROSINA_RTOF_SCI_OS_FIL_I ||Filament emission controller_Orthosource_Entrance1_current = Emission current [uA] (MC_Filament emission controller_Ortho_ENT1_I) ||measured ||std || ||2s ||1 min ||? +-15uA ||3 ||27 || || |- |54 ||NRNAR11D ||ROSINA_RTOF_SCI_SS_TRAP_V ||Storagesource trap voltage ||measured ||std || ||2s ||1 min ||? +-30V ||3 ||28 || || |- |56 ||NRNAR11E ||ROSINA_RTOF_SCI_HVVG_V ||Filament emission controller_voltage_~+24V (MC_Filament emission controller_HVVG_V) ||Flag On or off; measured ||std || ||2s ||1 min ||? 24+-6V ||3 ||29 || || |- |58 ||NRNAR11F ||ROSINA_RTOF_SCI_HEAT_VG_V ||Filament emission controller_voltage_~+24V (MC_Filament emission controller_HEAT_VG_V) ||Flag On or off; measured ||std || ||2s ||1 min ||? 24+-6V ||3 ||29 || || |- |60 ||NRNAR120 ||ROSINA_RTOF_SCI_FEC_TEMP ||Temperature of Filament emission controller ||measured ||std || ||2s ||1 min ||?<50 DegC ||3 ||29 || || |- |62 ||NRNAR121 ||ROSINA_RTOF_SCI_GCU_POW_1 ||GCU_Storagesource_Power On or Off (MC_GCU_POW_#1) ||measured ||std || ||200 ms ||1 min || ||4 || || || |- |64 ||NRNAR122 ||ROSINA_RTOF_SCI_GCU_POW_2 ||GCU_Orthosource_Power On or Off (MC_GCU_POW_#2) ||measured ||std || ||200 ms ||1 min || ||4 || || || |- |66 ||NRNAR123 ||ROSINA_RTOF_SCI_GCU_1_I ||Current in gas calibration unit 1 ||measured ||std ||GCU POWER ON ||200 ms ||1 min ||? ||2 ||12 || ||has to be 0 in standby On time maximal TBD |- |68 ||NRNAR124 ||ROSINA_RTOF_SCI_GCU_2_I ||Current in gas calibration unit 2 ||measured ||std ||GCU POWER ON ||200 ms ||1 min ||? ||2 ||12 || ||has to be 0 in standby On time maximal TBD |- |70 ||NRNAR125 ||ROSINA_RTOF_SCI_GCU_OUT_PRES_1 ||Output pressure of gas calibration unit 1 ||measured ||std ||GCU POWER ON ||200 ms ||1 min ||? ||2 ||12 ||yes || |- |72 ||NRNAR126 ||ROSINA_RTOF_SCI_GCU_OUT_PRES_2 ||Output pressure of gas calibration unit 2 ||measured ||std ||GCU POWER ON ||200 ms ||1 min ||? ||2 ||12 ||yes || |- |74 ||NRNAR127 ||ROSINA_RTOF_SCI_GCU_TEMP_PIRAN_1 ||Temperature of pirani in gas calibration unit 1 ||measured ||std || ||200 ms ||1 min ||?<50 DegC ||1 ||20 || || |- |76 ||NRNAR128 ||ROSINA_RTOF_SCI_GCU_TEMP_PIRAN_2 ||Temperature of pirani in gas calibration unit 2 ||measured ||std || ||200 ms ||1 min ||?<50 DegC ||1 ||20 || || |- |78 ||NRNAR129 ||ROSINA_RTOF_SCI_GEX_TEMP ||Gas Extraction Pulser_Temperature (MC_GEX_TEMP) ||measured ||std || ||2 s ||1 min ||?<50 DegC ||3 ||29 || || |- |80 ||NRNAR12A ||ROSINA_RTOF_SCI_HM_PW ||Hard mirror_Pulse width (MC_HM_PW) ||measured ||std || ||2 s ||1 min || ||4 || || || |- |82 ||NRNAR12B ||ROSINA_RTOF_SCI_HM_DEL ||hard mirror pulse delay ||measured ||std || ||2 s ||1 min || ||4 || || || |- |84 ||NRNAR12C ||ROSINA_RTOF_SCI_HM_TEMP ||hard mirror pulser temperature ||measured ||std || ||2 s ||1 min ||?<50 DegC ||3 ||29 || || |- |86 ||NRNAR12D ||ROSINA_RTOF_SCI_MOT_TORQUE ||Torque on cover opening mechanism motor ||measured ||std || ||2 s ||1 min ||? ||3 || || ||???? |- |88 ||NRNAR12E ||ROSINA_RTOF_SCI_MOT_TEMP ||Temperature on cover opening mechanism motor ||measured ||std || ||2 s ||1 min ||?<100 DegC ||3 || 29 || || |- |90 ||NRNAR12F || ||Cover opening mechanism Motor Position || ||std || ||2 s ||1 min || || || || || |- | || ||ROSINA_RTOF_SCI_MOT_ON_FLAG ||Motor On ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_MOT_POSITION_F ||Motor Positioning ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_VELOCITY_FLAG ||Motor Velocity ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_OPEN_SWITCH_F ||Open Switch ||Flag ||std || ||2 s ||1 min || ||2 ||13 || ||May only be reached in Search Switch Function |- | || ||ROSINA_RTOF_SCI_CLOSED_SWITCH_F ||Closed Switch ||Flag ||std || ||2 s ||1 min || ||2 ||13 || ||May only be reached in Search Switch Function |- | || ||ROSINA_RTOF_SCI_HALL_ERR_FLAG ||Hall sensor in motor Error ||Flag ||std || ||2 s ||1 min || ||3 || || || |- | || ||ROSINA_RTOF_SCI_HALL_ERR_FATAL_F ||Hall sensor Error Fatal ||Flag ||std || ||2 s ||1 min || ||2 ||14 || || |- | || ||ROSINA_RTOF_SCI_MOT_POSITION ||Motor Position ||Flag ||std || ||2 s ||1 min || ||2 ||13 || ||Measured ! FLAG ?? When Position exceeds range TBD ? |- |92 ||NRNAR130 || ||Main controller Power State 2 || ||std || ||2 s ||1 min || || || || || |- | || || || ||Flag ||std || ||2 s ||1 min || || || || || |- | || || || ||Flag ||std || ||2 s ||1 min || || || || || |- | || || || ||Flag ||std || ||2 s ||1 min || || || || || |- | || || || ||Flag ||std || ||2 s ||1 min || || || || || |- | || ||ROSINA_RTOF_SCI_HEATER_GAS_ON_F ||Heater Storage source On ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_HEATER_GAS_OFF_F ||Heater Storage source Off ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_HEATER_OS_ON_F ||Heater Ortho source On ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_HEATER_OS_OFF_F ||Heater Ortho source Off ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_MOT_HALL_ENA_F ||Motor Hall sensor Enable ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_MOT_HALL_DIS_F ||Motor Hall sensor Disable ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_MOT_DIR_OPEN_F ||Motor Direction Open ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_MOT_DIR_CLOSE_F ||Motor Direction Close ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_MOT_POW_ON_F ||Motor Power On ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_MOT_POW_OFF_F ||Motor Power Off ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_MOT_HIG_TOR_ON_F ||Motor High Torque On ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_MOT_HIG_TOR_OF_F ||Motor High Torque Off ||Flag ||std || ||2 s ||1 min || ||4 || || || |- |94 ||NRNAR131 || ||Main controller Pulser State || ||std || ||2 s ||1 min || || || || || |- | || ||ROSINA_RTOF_SCI_SS_PULS_ON_F ||Storage Pulser On ||flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_SS_PULS_OFF_F ||Storage Pulser Off ||flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_HM_PULS_ON_F ||Hard mirror Pulser On ||flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_HM_PULS_OFF_F ||Hard mirror Pulser Off ||flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_OS_PULS_ON_F ||Ortho Pulser On ||flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_OS_PULS_OFF_F ||Ortho Pulser Off ||flag ||std || ||2 s ||1 min || ||4 || || || |- | || || ||Spare ||flag ||std || ||2 s ||1 min || ||4 || || || |- |96 ||NRNAR132 || ||Main controller Power State 6 || ||std || ||2 s ||1 min || || || || || |- | || || ||Spare ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_DISABLE_ETS_LU_F ||Disable ETS Latch Up (Disable ETS LU) ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ENABLE_ETS_LU_F ||Enable ETS Latch Up (Enable ETS LU) ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETS_VCA_ON_FLAG ||Equivalent Time Sampling_analog voltage_On (ETS VCA On) ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETS_VCA_OFF_FLAG ||Equivalent Time Sampling_analog voltage_Off (ETS VCA Off) ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETS_VDD_ON_FLAG ||Equivalent Time Sampling_15V_On (ETS VDD On) ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETS_VDD_OFF_FLAG ||Equivalent Time Sampling_15V_Off (ETS VDD Off) ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETS_VCC_ON_FLAG ||Equivalent Time Sampling_5V_On (ETS VCC On) ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETS_VCC_OFF_FLAG ||Equivalent Time Sampling_5V_Off (ETS VCC Off) ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETS_VSH_VE_ON_F ||Equivalent Time Sampling_operation voltage_~58V_On (ETS VSH/VE On) ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETS_VSH_VE_OFF_F ||Equivalent Time Sampling_operation voltage_~58V_Off (ETS VSH/VE Off) ||Flag ||std || ||2 s ||1 min || ||4 || || || |- |98 ||NRNAR133 || ||Main controller Power State 8 || ||std || || ||1 min || || || || || |- | || ||ROSINA_RTOF_SCI_GCU1_ON_FLAG ||Gas calibration unit 1 On ||Flag ||std || ||200 ms ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_GCU1_OFF_FLAG ||Gas calibration unit 1 Off ||Flag ||std || ||200 ms ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_GCU2_ON_FLAG ||Gas calibration unit 2 On ||Flag ||std || ||200 ms ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_GCU2_OFF_FLAG ||Gas calibration unit 2 Off ||Flag ||std || ||200 ms ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_GCU1_VALVE_ON_F ||Gas calibration unit 1 Valve On ||Flag ||std || ||200 ms ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_GCU1_VALVE_OFF_F ||Gas calibration unit 1 Valve Off ||Flag ||std || ||200 ms ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_GCU2_VALVE_ON_F ||Gas calibration unit 2 Valve On ||Flag ||std || ||200 ms ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_GCU2_VALVE_OF_F ||Gas calibration unit 2 Valve Off ||Flag ||std || ||200 ms ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_FEC_VCC_ON_FLAG ||Filament emission controller_5V_On (Filament emission controller VCC On) ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_FEC_VCC_OFF_FLAG ||Filament emission controller_5V_Off (Filament emission controller VCC Off) ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_FEC_VDD_ON_FLAG ||Filament emission controller_15V_On (Filament emission controller VDD On) ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_FEC_VDD_OFF_FLAG ||Filament emission controller_15V_Off (Filament emission controller VDD Off) ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_FEC_HEAT_VG_ON_F ||Filament emission controller_24V_On (Filament emission controller Heater VG On) ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_FEC_HEAT_VG_OF_F ||Filament emission controller_24V_Off (Filament emission controller Heater VG Off) ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_FEC_HV_VG_ON_F ||Filament emission controller_24V_On (Filament emission controller HV VG On) ||Flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_FEC_HV_VG_OF_F ||Filament emission controller_24V_Off (Filament emission controller HV VG Off) ||Flag ||std || ||2 s ||1 min || ||4 || || || |- |100 ||NRNAR134 ||ROSINA_RTOF_SCI_MC_ETSL_TEMP ||Main controller_WCS_TEMP = ETSL_TEMP ||measured ||std || ||2 s ||1 min ||?<50 DegC ||3 ||17 || ||NEW: ETSLIGHT_TEMP |- |102 ||NRNAR135 ||ROSINA_RTOF_SCI_MC_ETS_TEMP_CLK ||ETS temperature on clock circuitry ||measured ||std || ||2 s ||1 min || ||3 ||15 || ||Max TemperatureTBD |- |104 ||NRNAR136 ||ROSINA_RTOF_SCI_MC_ETS_TEMP_DIG ||ETS temperature on digital circuitry ||measured ||std || ||2 s ||1 min ||?<50 DegC ||3 ||15 || ||Max TemperatureTBD |- |106 ||NRNAR137 ||ROSINA_RTOF_SCI_PSDC_ELB_OS ||Entrance Lens_B_Orthosource_voltage=lower lens (PSDC_ELB_I) ||measured ||std || ||2 s ||1 min ||?+-5V ||3 ||28 || || |- |108 ||NRNAR138 ||ROSINA_RTOF_SCI_PSDC_ELA_OS ||Entrance Lens_A_Orthosource_voltage=upper lens (PSDC_ELA_I) ||measured ||std || ||2 s ||1 min ||?+-5V ||3 ||28 || || |- |110 ||NRNAR139 ||ROSINA_RTOF_SCI_PSDC_GR_OS ||Extraction Grid_Orthosource_voltage (PSDC_GR_I) ||measured ||std || ||2 s ||1 min ||?+-5V ||3 ||28 || || |- |112 ||NRNAR13A ||ROSINA_RTOF_SCI_PSDC_BP_OS ||Backplane_Orthosource_voltage (PSDC_BP_I) ||measured ||std || ||2 s ||1 min ||?+-5V ||3 ||28 || || |- |114 ||NRNAR13B ||ROSINA_RTOF_SCI_PSDC_GR_SS ||Extraction Grid_Storagesource_voltage (PSDC_GR_G) ||measured ||std || ||2 s ||1 min ||?+-5V ||3 ||28 || || |- |116 ||NRNAR13C ||ROSINA_RTOF_SCI_PSDC_BP_SS ||Backplane_Storagesource_voltage (PSDC_BP_G) ||measured ||std || ||2 s ||1 min ||?+-5V ||3 ||28 || || |- |118 ||NRNAR13D ||ROSINA_RTOF_SCI_HV1_SL_SS ||high voltage Storage Source lense ||measured ||std || ||2 s ||1 min ||?+-200V || ||28 || || |- |120 ||NRNAR13E ||ROSINA_RTOF_SCI_HV1_A2_SS ||Accerleration_2_Storagesource_voltage (high voltage1_A2_G) ||measured ||std || ||2 s ||1 min ||?+-200V ||3 ||28 || || |- |122 ||NRNAR13F ||ROSINA_RTOF_SCI_HV1_A1_SS ||Acceleration_1_Storagesource_voltage (high voltage1_A1_G) ||measured ||std || ||2 s ||1 min ||?+-100V ||3 ||28 || || |- |124 ||NRNAR140 ||ROSINA_RTOF_SCI_HV1_SL_OS ||high voltage on ortho source lense ||measured ||std || ||2 s ||1 min ||?+-200V || ||28 || || |- |126 ||NRNAR141 ||ROSINA_RTOF_SCI_HV1_A2_OS ||Acceleration_2_Orthosource_voltage (high voltage1_A2_I) ||measured ||std || ||2 s ||1 min ||?+-200V ||3 ||28 || || |- |128 ||NRNAR142 ||ROSINA_RTOF_SCI_HV1_A1_OS ||Accleration_1_Orthosource_voltage (high voltage1_A1_I) ||measured ||std || ||2 s ||1 min ||?+-100V ||3 ||28 || || |- |130 ||NRNAR143 ||ROSINA_RTOF_SCI_HV2_P_SS ||high voltage on storage source pulser ||measured ||std || ||2 s ||1 min ||?+-50V ||3 ||28 || || |- |132 ||NRNAR144 ||ROSINA_RTOF_SCI_HV2_P_OS ||high voltage on ortho source pulser ||measured ||std || ||2 s ||1 min ||?+-50V ||3 ||28 || || |- |134 ||NRNAR145 ||ROSINA_RTOF_SCI_HV1_D ||Drift_voltage (high voltage1_D) ||measured ||std || ||2 s ||1 min ||?+-100V ||3 ||28 || || |- |136 ||NRNAR146 ||ROSINA_RTOF_SCI_HV2_HM3 ||high voltage Hard mirror3 ||measured ||std || ||2 s ||1 min ||?+-50V ||3 ||28 || ||PSU_HM3 |- |138 ||NRNAR147 ||ROSINA_RTOF_SCI_HV1_R1 ||high voltage reflectron 1 ||measured ||std || ||2 s ||1 min ||?+-50V ||3 ||28 || ||PSU_R1 |- |140 ||NRNAR148 ||ROSINA_RTOF_SCI_HV1_R2 ||high voltage reflectron 2 ||measured ||std || ||2 s ||1 min ||?+-200V ||3 ||28 || ||PSU_R2 |- |142 ||NRNAR149 ||ROSINA_RTOF_SCI_HV1_RL ||high voltage on reflectron lense ||measured ||std || ||2 s ||1 min ||?+-200V ||3 ||28 || ||PSU_RL |- |144 ||NRNAR14A ||ROSINA_RTOF_SCI_HV1_HM1 ||high voltage Hard mirror1 ||measured ||std || ||2 s ||1 min ||?+-100V ||3 ||28 || ||PSU_HM1 |- |146 ||NRNAR14B ||ROSINA_RTOF_SCI_HV2_HM2 ||high voltage Hard mirror2 ||measured ||std || ||2 s ||1 min ||?+-50V ||3 ||28 || ||PSU_HM2 |- |148 ||NRNAR14C ||ROSINA_RTOF_SCI_HV1_HML ||high voltage Hard mirror lense ||measured ||std || ||2 s ||1 min ||?+-200V ||3 ||28 || ||PSU_HML |- |150 ||NRNAR14D ||ROSINA_RTOF_SCI_HV2_M_OS ||high voltage on ortho source detector ||measured ||std || ||2 s ||1 min ||?+-150V ||2 ||18 || ||M_I |- |152 ||NRNAR14E ||ROSINA_RTOF_SCI_HV2_M_SS ||high voltage on storage source detector ||measured ||std || ||2 s ||1 min ||?+-150V ||2 ||18 || ||M_G |- |154 ||NRNAR14F ||ROSINA_RTOF_SCI_PSDC_E2_OS ||Entrance2_Orthosource_voltage (PSDC_E2_I) ||measured ||std || ||2 s ||1 min ||?+-5V ||3 ||28 || || |- |156 ||NRNAR150 ||ROSINA_RTOF_SCI_PSDC_TEMP_BP_OS ||Temperature_Backplane_Orthosource (PSDC_Temp_BP_I) ||measured ||std || ||2 s ||1 min || ||4 || || || |- |158 ||NRNAR151 ||ROSINA_RTOF_SCI_PSDC_TEMP_BP_SS ||Temperature_Backplane_Storagesource (PSDC_Temp_BP_G) ||measured ||std || ||2 s ||1 min || ||4 || || || |- |160 ||NRNAR152 ||ROSINA_RTOF_SCI_PSU_TEMP_MCP_OS ||Temperature_Multi channel plates_Orthosource (PSU_Temp_MCP_I) ||measured ||std || ||2 s ||1 min ||?<50 DegC ||3 ||30 || || |- |162 ||NRNAR153 ||ROSINA_RTOF_SCI_SPARE2_FLAG ||Temperature_Multi channel plates_Storagesource (PSU_Temp_MCP_G) ||measured ||std || ||2 s ||1 min || ||4 || || || |- |164 ||NRNAR154 ||ROSINA_RTOF_SCI_PSU_TEMP_HV1 ||Temperature of high voltage board 1 ||measured ||std || ||2 s ||1 min ||?<50 DegC ||3 ||19 || || |- |166 ||NRNAR155 ||ROSINA_RTOF_SCI_PSU_TEMP_LVPS ||Temperature of low voltage board ||measured ||std || ||2 s ||1 min ||?<50 DegC ||3 ||19 || || |- |168 ||NRNAR156 || || || ||std || ||2 s ||1 min || || || || ||Not used -- ETS LIGHT version |- | || || || || ||std || ||2 s ||1 min || || || || || |- | || || || || ||std || ||2 s ||1 min || || || || || |- | || || || || ||std || ||2 s ||1 min || || || || || |- | || || || || ||std || ||2 s ||1 min || || || || || |- | || || || || ||std || ||2 s ||1 min || || || || || |- | || || || || ||std || ||2 s ||1 min || || || || || |- | || || || || ||std || ||2 s ||1 min || || || || || |- |169 || || || || ||std || ||2 s ||1 min || || || || || |- |170 ||NRNAR157 || ||ETS Status 1 || ||std || ||2 s ||1 min || || || || || |- | || ||ROSINA_RTOF_SCI_ETSL_OSP_STAT_F ||ETS Ortho Pulser Status ||flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETS_TRIG_F ||ETS Storage Pulser Status ||flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETSL_SYNC_STAT_F ||ETS Synchronization Status ||flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETSL_CAL_TRIG_F ||ETS Trigger Status ||flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETSL_DAT_READ_F ||ETS Data Readout Status ||flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETSL_ACQ_STAT_F ||ETS Acquisition Status ||flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETSL_DTS_CANC_F ||ETS Delayed time sampling Status Cancel ||flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETSL_DTS_STAT_F ||ETS Delayed time sampling Status ||flag ||std || ||2 s ||1 min || ||4 || || || |- |171 || || ||ETS Status 2 || ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETSL_INPUT_F ||ETS Input Status ||flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETSL_CAL_POW_F ||ETS Cal. Power Status ||flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETSL_ADC_HIRM_F ||ETS ADC High resolution mode Status ||flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETSL_RAM_THR_F ||ETS RAM Threshold ||Value ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETSL_FIFO_THR_F ||ETS FIFO Threshold ||Value ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETSL_LU_ENA_F ||ETS Latchup Enabled ||flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETSL_LU_DETECT_F ||ETS Latchup Detected ||flag ||std || ||2 s ||1 min || ||3 ||16 || || |- |172 ||NRNAR158 || ||ETS Status 3 || ||std || ||2 s ||1 min || || || || || |- | || ||ROSINA_RTOF_SCI_ETSL_ADC_POW_F ||ETS ADC Power Status ||flag ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETSL_ADC_THR_F ||ETS ADC Threshold ||Value ||std || ||2 s ||1 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETSL_ML_MODE_F ||ETS mass line mode Mode ||Value ||std || ||2 s ||1 min || ||4 || || || |- | || || ||Spare || ||std || ||2 s ||1 min || || || || || |- |173 || ||ROSINA_RTOF_SCI_ETSL_THR_LEVEL_F ||ETS Threshold Level ||Value ||std || ||2 s ||1 min || ||4 || || || |- |174 ||NRNAR159 ||ROSINA_RTOF_SCI_ETSL_EX_DEL ||ETS Extraction Delay ||Value ||std || ||2 s ||1 min || ||4 || || || |- |176 ||NRNAR15A ||ROSINA_RTOF_SCI_ETSL_TOF ||ETS Time of flight ||Value ||std || ||2 s ||1 min || ||4 || || || |- |178 ||NRNAR15B ||ROSINA_RTOF_SCI_ETSL_CAL_DEL ||ETS Cal. Start Delay ||Value ||std || ||2 s ||1 min || ||4 || || || |- |180 ||NRNAR15C ||ROSINA_RTOF_SCI_ETSL_CAL_PH ||ETS Cal. Pulse Height ||Value ||std || ||2 s ||1 min || ||4 || || || |- |182 ||NRNAR15D ||ROSINA_RTOF_SCI_ETSL_CAL_PW ||ETS Cal. Pulse Width ||Value ||std || ||2 s ||1 min || ||4 || || || |- |184 ||NRNAR15E || ||Spare || ||std || ||2 s ||1 min || || || || || |} {| TABLE !Pos !!RSDB !! !!Name !! !!ext !! !! !! !! !! !! !! !! ! |- |0 ||NRNAR160 || ||RTOF Ext. HK Header || ||ext || ||2 s ||5 min || || || || || |- |1 || || ||Spare || ||ext || ||2 s ||5 min || || || || || |- |2 ||NRNAR161 ||ROSINA_RTOF_SCI_MC_FEC_PVCC ||Filament emission controller_Plus VCC_voltage (Filament emission controller_PVCC_V) ||measured ||ext || ||2 s ||5 min ||? ||3 ||21 || ||FEC Supply out of range |- |4 ||NRNAR162 ||ROSINA_RTOF_SCI_MC_FEC_MVCC ||Filament esmission controller_Minus VCC_voltage (Filament emission controller_MVCC_V) ||measured ||ext || ||2 s ||5 min ||? ||3 ||21 || ||FEC Supply out of range |- |6 ||NRNAR163 ||ROSINA_RTOF_SCI_MC_FEC_PVDD ||Filament emission controller_Plus VDD_voltage (Filament emission controller_PVDD_V) ||measured ||ext || ||2 s ||5 min ||? ||3 ||21 || ||FEC Supply out of range |- |8 ||NRNAR164 ||ROSINA_RTOF_SCI_MC_FEC_MVDD ||Filament emission controller_Minus VDD_voltage(Filament emission controller_MVDD_V) ||measured ||ext || ||2 s ||5 min ||? ||3 ||21 || ||FEC Supply out of range |- |10 ||NRNAR165 ||ROSINA_RTOF_SCI_MC_GCU_T_TANK1 ||Temperature on gas calibration unit tank 1 ||measured ||ext ||GCU Power On ||2 s ||5 min ||> 50 DegC ||3 ||20 || || |- |12 ||NRNAR166 ||ROSINA_RTOF_SCI_MC_GCU_T_TANK2 ||Temperature on gas calibration unit tank 2 ||measured ||ext ||GCU Power On ||2 s ||5 min ||> 50 DegC ||3 ||20 || || |- |14 ||NRNAR167 ||ROSINA_RTOF_SCI_MC_GCU_PR_IN1 ||Gas calibration unit pressure 1 on tank side ||measured ||ext || ||2 s ||5 min || ||4 || || || |- |16 ||NRNAR168 ||ROSINA_RTOF_SCI_MC_GCU_PR_IN2 ||Gas calibration unit pressure 2 on tank side ||measured ||ext || ||2 s ||5 min || ||4 || || || |- |18 ||NRNAR169 ||ROSINA_RTOF_SCI_MC_GEX_PVCC ||Gas Extraction Pulser_+5V (Main controller_GEX_PVCC) ||measured ||ext ||Pulser Power On ||2 s ||5 min ||? ||3 ||22 || ||Pulser Supply out of range |- |20 ||NRNAR16A ||ROSINA_RTOF_SCI_MC_GEX_VD ||Gas Extraction Pulser_28V (Main controller_GEX_VD) ||measured ||ext ||VD ON ||2 s ||5 min ||? ||3 ||22 || ||Pulser Supply out of range |- |22 ||NRNAR16B ||ROSINA_RTOF_SCI_MC_IEX_PVCC ||Ion Extraction Pulser_+5V (Main controller_IEX_PVCC) ||measured ||ext || ||2 s ||5 min ||? ||3 ||22 || ||Pulser Supply out of range |- |24 ||NRNAR16C ||ROSINA_RTOF_SCI_MC_IEX_VD ||Ion Extraction Pulser_28V (Main controller_IEX_VD) ||measured ||ext || ||2 s ||5 min ||? ||3 ||22 || ||Pulser Supply out of range |- |26 ||NRNAR16D ||ROSINA_RTOF_SCI_MC_HM_PVCA ||Hard Mirror_+5V analog (Main controller_Hard mirror_PVCA) ||measured ||ext || ||2 s ||5 min ||? ||3 ||22 || ||Pulser Supply out of range |- |28 ||NRNAR16E ||ROSINA_RTOF_SCI_MC_HM_VD ||Hard Mirror_28V (Main controller_Hard mirror_VD) ||measured ||ext || ||2 s ||5 min ||? ||3 ||22 || ||Pulser Supply out of range |- |30 ||NRNAR16F ||ROSINA_RTOF_SCI_MC_ETSL_PVCC || = ETSLIGHT PVCA || ||ext || ||2 s ||5 min ||? ||3 ||23 || || |- |32 ||NRNAR170 ||ROSINA_RTOF_SCI_MC_ETSL_MVCA || = ETSLIGHT MVCA || ||ext || ||2 s ||5 min ||? ||3 ||23 || || |- |34 ||NRNAR171 ||ROSINA_RTOF_SCI_MC_ETSL_PVDD || = ETSLIGHT PVDD || ||ext || ||2 s ||5 min ||? ||3 ||23 || || |- |36 ||NRNAR172 ||ROSINA_RTOF_SCI_MC_ETSL_MVDD || = ETSLIGHT MVDD || ||ext || ||2 s ||5 min ||? ||3 ||23 || || |- |38 ||NRNAR173 ||ROSINA_RTOF_SCI_MC_MOT_PVCC || = ETSLIGHT PVCC || ||ext || ||2 s ||5 min ||? ||3 ||23 || || |- |40 ||NRNAR174 ||ROSINA_RTOF_SCI_MC_OMC_SW_OPEN ||Motor_Switch_open position (Main controller_OMain controller_SW_OPEN) ||measured ||ext ||Motor Power On ||2 s ||5 min ||? ||2 ||13 || || |- |42 ||NRNAR175 ||ROSINA_RTOF_SCI_MC_OMC_SW_CLOSED ||Motor_Switch_close position (Main controller_OMain controller_SW_CLOSED) ||measured ||ext ||Motor Power On ||2 s ||5 min ||? ||2 ||13 || || |- |44 ||NRNAR176 ||ROSINA_RTOF_SCI_MC_HEAT_ION_V ||Heater_Orthosource_24V (Main controller_HEAT_Ortho_VG) ||measured ||ext ||FEC Heater Power On ||2 s ||5 min ||? ||3 ||24 || ||FEC_HEAT_VG Heater Supply out of range |- |46 ||NRNAR177 ||ROSINA_RTOF_SCI_MC_HEAT_GAS_V ||Heater_Storagesource_24V (Main controller_HEAT_Storage _VG) ||measured ||ext || ||2 s ||5 min ||? ||3 ||24 || ||FEC_HEAT_VG Heater Supply out of range |- |48 ||NRNAR178 ||ROSINA_RTOF_SCI_MC_ETS_33V ||Main controller_ETS_33V ||measured ||ext || ||2 s ||5 min ||? ||3 ||25 || ||ETS Supply out of range |- |50 ||NRNAR179 ||ROSINA_RTOF_SCI_MC_ETS_PVCA ||Equivalent Time Sampling_Plus VCA_voltage (Main controller_ETS_PVCA) ||measured ||ext ||ETS is powered ||2 s ||5 min ||? ||3 ||25 || ||ETS Supply out of range |- |52 ||NRNAR17A ||ROSINA_RTOF_SCI_MC_ETS_MVCA ||Equivalent Time Sampling_Minus VCA_voltage (Main controller_ETS_MVCA) ||measured ||ext ||ETS is powered ||2 s ||5 min ||? ||3 ||25 || ||ETS Supply out of range |- |54 ||NRNAR17B ||ROSINA_RTOF_SCI_MC_ETS_VE ||Equivalent Time Sampling_8V (Main controller_ETS_VE) ||measured ||ext ||NEVER ! ||2 s ||5 min || || || || ||Not used anymore !!!! |- |56 ||NRNAR17C ||ROSINA_RTOF_SCI_MC_ETS_PVDD ||Equivalent Time Sampling_Plus VDD_voltage (Main controller_ETS_PVDD) ||measured ||ext ||ETS is powered ||2 s ||5 min ||? ||3 ||25 || ||ETS Supply out of range |- |58 ||NRNAR17D ||ROSINA_RTOF_SCI_MC_ETS_MVDD ||Equivalent Time Sampling_Minus VDD_voltage (Main controller_ETS_MVDD) ||measured ||ext ||ETS is powered ||2 s ||5 min ||? ||3 ||25 || ||ETS Supply out of range |- |60 ||NRNAR17E ||ROSINA_RTOF_SCI_MC_ETS_VSH ||Equivalent Time Sampling_operation voltage for Analog Digital Converter_~58V (Main controller_ETS_VSH) ||measured ||ext ||ETS ADC section is powered ||2 s ||5 min ||? ||3 ||25 || ||ETS Supply out of range |- |62 ||NRNAR17F ||ROSINA_RTOF_SCI_MC_ETSL_PVCA ||Main controller_Spare --> ETSLIGHT_PVCA ||measured ||ext ||ETS LIGHT is powered ||2 s ||5 min ||? ||3 ||26 || ||ETSLIGHT PVCA ETS LIGHT supply out of range |- |64 ||NRNAR180 ||? ||Main controller_Spare || ||ext || ||2 s ||5 min || || || || || |- |66 ||NRNAR181 ||ROSINA_RTOF_SCI_PSU_P5V_V ||Power supply unit +5_Volt value ||measured ||ext || ||2 s ||5 min ||? ||1 || || || |- |68 ||NRNAR182 ||ROSINA_RTOF_SCI_PSU_M5V_V ||Power supply unit -5_Volt value ||measured ||ext || ||2 s ||5 min ||? ||1 || || || |- |70 ||NRNAR183 ||ROSINA_RTOF_SCI_PSU_P15V_V ||Power supply unit +15_Volt value ||measured ||ext || ||2 s ||5 min ||? ||1 || || || |- |72 ||NRNAR184 ||ROSINA_RTOF_SCI_PSU_M15V_V ||Power supply unit -15_Volt value ||measured ||ext || ||2 s ||5 min ||? ||1 || || || |- |74 ||NRNAR185 ||ROSINA_RTOF_SCI_PSU_24V_V ||Power supply unit +24_Volt value ||measured ||ext || ||2 s ||5 min ||? ||1 || || || |- |76 ||NRNAR186 ||ROSINA_RTOF_SCI_PSU_8V_V ||Power supply unit +8_Volt value ||measured ||ext || ||2 s ||5 min ||? ||4 || || ||Not used anymore by ETS |- |78 ||NRNAR187 ||ROSINA_RTOF_SCI_PSU_5VA_V ||Power supply unit +5_Add_Volt value ||measured ||ext || ||2 s ||5 min ||? ||1 || || || |- |80 ||NRNAR188 ||ROSINA_RTOF_SCI_PSU_40V_V ||Power supply unit +40_Volt value ||measured ||ext || ||2 s ||5 min ||? ||1 || || || |- |82 ||NRNAR189 ||ROSINA_RTOF_SCI_PSU_70V_V ||Power supply unit +70_Volt value ||measured ||ext || ||2 s ||5 min ||? ||1 || || || |- |84 ||NRNAR18A ||ROSINA_RTOF_SCI_PSU_P5V_I ||Power supply unit +5_Cur ||measured ||ext || ||2 s ||5 min || ||4 || || || |- |86 ||NRNAR18B ||ROSINA_RTOF_SCI_PSU_M5V_I ||Power supply unit -5_Cur ||measured ||ext || ||2 s ||5 min || ||4 || || || |- |88 ||NRNAR18C ||ROSINA_RTOF_SCI_PSU_P15V_I ||Power supply unit +15_Cur ||measured ||ext || ||2 s ||5 min ||>500 mA ||1 || || || |- |90 ||NRNAR18D ||ROSINA_RTOF_SCI_PSU_M15V_I ||Power supply unit -15_Cur ||measured ||ext || ||2 s ||5 min ||>300 mA ||1 || || || |- |92 ||NRNAR18E ||ROSINA_RTOF_SCI_PSU_24V_I ||Power supply unit +24_Cur ||measured ||ext || ||2 s ||5 min || ||4 || || ||checked for sudden increase |- |94 ||NRNAR18F ||ROSINA_RTOF_SCI_PSU_40_70_I ||Power supply unit +40/70_Cur ||measured ||ext || ||2 s ||5 min || ||4 || || || |- |96 ||NRNAR190 ||ROSINA_RTOF_SCI_PSU_5VA_I ||Power supply unit_+5V_Additional_current ||measured ||ext || ||2 s ||5 min || ||4 || || || |- |98 ||NRNAR191 ||ROSINA_RTOF_SCI_PSU_8V_I ||Power supply unit +8_Cur ||measured ||ext || ||2 s ||5 min || ||4 || || ||Not used by any board |- |100 ||NRNAR192 || || || ||ext || ||2 s ||5 min || || || || || |- |104 ||NRNAR193 ||ROSINA_RTOF_SCI_ETS_NOE_HIGH_V2 ||Postprocessor (ETS number of extractions High) ||value ||ext || ||2 s ||5 min || ||4 || || || |- | || || ||Postprocessor (number of extractions High value) ||value ||ext || ||2 s ||5 min || ||4 || || || |- |106 ||NRNAR194 ||ROSINA_RTOF_SCI_ETS_NOE_LOW_V2 ||Postprocessor (ETS number of extractions Low) ||value ||ext || ||2 s ||5 min || ||4 || || || |- | || || ||Postprocessor (number of extractions Low value) ||value ||ext || ||2 s ||5 min || ||4 || || || |- | || ||ROSINA_RTOF_SCI_ETS_NOE_STAT2_F ||Postprocessor (number of extractions Status) ||value ||ext || ||2 s ||5 min || ||4 || || || |- |108 ||NRNAR195 || ||Spare 1 || ||ext || ||2 s ||5 min || || || || || |- |110 ||NRNAR196 || ||Spare 2 || ||ext || ||2 s ||5 min || || || || || |} {| TABLE !Action !!Description ! |- |1 ||Turn RTOF off, turn RTOF on again, return to current sensor state. If problem still exists, turn RTOF completely off until appropriate reaction by operator. |- |2 ||Turn RTOF off, turn RTOF on again, repeat current measurement mode. If problem still exists, turn RTOF completely off until appropriate reaction by operator. |- |3 ||Turn off filament, try it again. If problem still exists, turn RTOF completely off until appropriate reaction by operator. |- |4 ||Stop cover operation, TBD further action. |- |5 ||Stop GCU operation, TBD further action. |- |6 ||Turn RTOF off, TBD further action. |- |7 ||Stop GCU operation, if problems still exists, turn RTOF completely off, TBD further action. |- |8 ||Stop RDP heater operation, try it again. If problems still exists, TBD further action. |- |9 ||Stop IS heater operation, try it again. If problems still exists, TBD further action. |- |10 ||Stop GCU operation, try it again. If problems still exists, TBD further action. |} {| TABLE !Action !!Description ! |- |11 ||Decrease Filament Heater level fast until the 24V current is in a nominal range |- |12 ||Switch off GCU valve |- |13 ||Stop Motor by switching motor power off |- |14 ||Stop Motor and try to drive it without hall sensor |- |15 ||ETS CLOCKBOARD OVERTEMP: Switch off ETS |- |16 ||ETS LATCHUP DETECTED : TBD |- |17 ||ETS LIGHT OverTemperature : TBD |- |18 ||MCP Problem - Decrease MCP voltage TBD |- |19 ||Overtemperature on PSU Boards : TBD |- |20 ||GCU Overtemperature : Possibility to loose gas through valve |- |21 ||FEC Supply out of range |- |22 ||Pulser Supply out of range (used for Gas/Ion/HM-Pulser |- |23 ||ETS Light Supply out of range |- |24 ||Fec Heater Supply voltage out of range : Power down Heater by setting MC to 0 and switching Heaters off |- |25 ||ETS Supply out of range Power down ETS |- |26 ||ETS LIGHT Supply out of range Power down ETS LIGHT |- |27 ||Switch off filament |- |28 ||Switch off corresponding voltage (do not switch off all or the 9kV) |- |29 ||Switch off corresponding unit (FEC / Pulser, GCU) |- |30 ||Switch off both detectors |- | || |- | ||8V: Monitoring? |} COPS HK monitoring The converted text has been copied to the clip-board : Just copy (Ctrl-V) in the ASCII document. Good luck ! {| TABLE !RSDB !!PDS Keyword !!HK-Mnemonic old !!Description !!HK Type !!Std./Ext. Pack. !!Monitoring if !!Sampling frequency !!Transmission frequency S/C !!Range !!Criti-cality !!Action, if outside range !!Used for regulation !!Remarks ! |- |NRNAC101 || ||COPS Std. HK Header || || || || || || || || || || || |- | || ||Range Status || || || || || || || || || || || |- | ||ROSINA_COPS_COMMAND_ERROR_FLAG ||Command Error ||COPS command Error ||Flag ||Std. ||Measurement ||2s ||1min ||On / Off ||TBD || || || |- | || ||Spare || || || || || || || || || || || |- | ||ROSINA_COPS_FIL_ION_R_FLAG ||Filament Ion Range ||Config of the electrometer for the Ion current in Nude Gauge ||Flag ||Std. ||Measurement with Nude Gauge ||2s ||1min ||Low / Medium / High ||4 ||no action || || |- | ||ROSINA_COPS_MIC_ION_R_FLAG ||Microtip Ion Range ||Config of the electrometer for the Ion current in Ram Gauge ||Flag ||Std. ||Measurement with Ram Gauge ||2s ||1min ||Low / Medium / High ||4 ||no action || || |- | ||ROSINA_COPS_FIL_EMISSION_R_FLAG ||Filament Imission Range ||Config for the Immision current measurement in Nude Gauge ||Flag ||Std. ||Measurement with Nude Gauge ||2s ||1min ||Low / High ||4 ||no action || || |- | ||ROSINA_COPS_MIC_EMISSION_R_FLAG ||Microtip Imission Range ||Config for the Immision current measurement in Ram Gauge ||Flag ||Std. ||Measurement with Ram Gauge ||2s ||1min ||Low / High ||4 ||no action || || |- |NRNAC102 ||ROSINA_COPS_HK_COUNTER ||COPS HK counter ||Housekeeping counter ||counter ||Std. ||Measurement ||2s ||1min ||not applicable ||TBD || || || |- |NRNAC103 ||ROSINA_COPS_CMD_COUNTER ||COPS Cmd counter ||Command counter ||counter ||Std. ||Measurement ||2s ||1min ||not applicable ||TBD || || || |- |NRNAC104 ||ROSINA_COPS_CMD_ERROR_COUNTER ||COPS Cmd Error counter ||Command error counter ||counter ||Std. ||Measurement ||2s ||1min ||not applicable ||TBD || || || |- |NRNAC105 ||ROSINA_COPS_CMD_ERROR_POSITION ||COPS Cmd Error position || || || || || || || || || || || |- |NRNAC106 ||ROSINA_COPS_HK_ERROR_COUNTER ||COPS HK Error counter ||Housekeeping errors counter ||counter ||Std. ||Measurement ||2s ||1min ||not applicable ||TBD || || || |- |NRNAC107 ||ROSINA_COPS_MODE ||COPS S/W mode || || || || || || || || || || || |- |NRNAC108 ||ROSINA_COPS_STATUS ||COPS S/W status || || || || || || || || || || || |- |NRNAC109 ||ROSINA_COPS_PRESSURE_NG ||Pressure NG ||Pressure in the Nude Gauge ||Calculated ||Std. ||Measurement with Nude Gauge ||2s ||1min ||<1E-4 / <1E-5 (TBC) ||1 / 2 || || || |- |NRNAC10A ||ROSINA_COPS_PRESSURE_RG ||Pressure RG ||Pressure in the Ram Gauge ||Calculated ||Std. ||Measurement with Ram Gauge ||2s ||1min ||<1E-4 / <1E-5 (TBC) ||1 / 2 || || || |- |NRNAC10B ||ROSINA_COPS_CALIB_FAC_OFFSET_NG ||Calibration factor NG ||Value entering in the calculation of the Nude Gauge pressure ||Fixed value ||Std. ||Measurement with Nude Gauge ||2s ||1min ||not applicable ||4 ||no action || || |- |NRNAC10C ||ROSINA_COPS_CALIB_FAC_OFFSET_RG ||Calibration factor RG ||Value entering in the calculation of the Ram Gauge pressure ||Fixed value ||Std. ||Measurement with Ram Gauge ||2s ||1min ||not applicable ||4 ||no action || || |- |NRNAC10D || ||Active Filament/Microtips ||Filament or/and Microtips activated ||Flag ||Std. ||Measurement ||2s ||1min ||Filament / Micotips ||4 ||no action || || |- | || ||Spare || || || || || || || || || || || |- | ||ROSINA_COPS_MIC_ARRAY1_FLAG ROSINA_COPS_MIC_ARRAY2_FLAG ROSINA_COPS_MIC_ARRAY3_FLAG ROSINA_COPS_MIC_ARRAY4_FLAG ROSINA_COPS_MIC_ARRAY5_FLAG ROSINA_COPS_MIC_ARRAY6_FLAG ROSINA_COPS_MIC_ARRAY7_FLAG ROSINA_COPS_MIC_ARRAY8_FLAG ||Microtips Array ||Microtips configuration, shows wich Microtip groups are enabled or not ||Flag array ||Std. ||Measurement with Ram Gauge ||2s ||1min ||1 to 8 Micotips On /Off ||4 ||no action || || |- | || ||Spare || || ||Std. || || || || || || || || |- | ||ROSINA_COPS_FILAMENT_FLAG ||Filament ||Filament configuration, left or right filament enabled ||Flag ||Std. ||Measurement with Nude Gauge ||2s ||1min ||Left / Right ||4 ||no action || || |- |NRNAC10E || ||Spare || || || || || || || || || || || |- |NRNAC10F ||ROSINA_COPS_ION_I_FILAMENT ||Ion I Filament ||Ion current in the Nude Gauge ||Measured ||Std. ||Measurement with Nude Gauge ||2s ||1min ||> 0pA ||3 ||3 || || |- | || || || || || || || || || || || || || |- |NRNAC110 ||ROSINA_COPS_ION_I_MICROTIP ||Ion I Microtip ||Ion current in the Ram Gauge ||Measured ||Std. ||Measurement with Ram Gauge ||2s ||1min ||> 0pA ||3 ||4 || || |- | || || || || || || || || || || || || || |- |NRNAC111 ||ROSINA_COPS_EMISSION_I_FILAMENT ||Imission I Filament ||Trapped current at Nude Gauge grid ||Measured ||Std. ||Measurement with Nude Gauge ||2s ||1min ||< 1000uA depending on mode (high) ||3 ||5 || || |- | || || || || || || || || || || || || || |- |NRNAC112 ||ROSINA_COPS_EMISSION_I_MICROTIP ||Imission I Microtip ||Trapped current at Ram Gauge grid ||Measured ||Std. ||Measurement with Ram Gauge ||2s ||1min ||< 1000uA depending on mode (high) ||3 ||5 || || |- | || || || || || || || || || || || || || |- |NRNAC113 ||ROSINA_COPS_ANODE_GRID_V_FIL ||AnodeGrid V Filament ||Voltage applied to the grid of the Nude Gauge ||Measured ||Std. ||Measurement with Nude Gauge ||2s ||1min ||Nominal value +/- 5% ||2 ||5 || || |- |NRNAC115 ||ROSINA_COPS_ANODE_GRID_V_MIC ||AnodeGrid V Microtip ||Voltage applied to the grid of the Ram Gauge ||Measured ||Std. ||Measurement with Ram Gauge ||2s ||1min ||Nominal value +/- 5% ||2 ||5 || || |- |NRNAC116 ||ROSINA_COPS_FILAMENT_I ||Filament I ||Current in the selected filament ||Measured ||Std. ||Measurement with Nude Gauge ||2s ||1min ||< 1A ||2 ||5 || || |- |NRNAC117 ||ROSINA_COPS_EXTRACTION_V_MIC ||Extraction V Microtip ||Extraction voltage applied for mtips in the Ram Gauge ||Measured ||Std. ||Measurement with Ram Gauge ||2s ||1min ||< 80V ||2 ||6 || || |- |NRNAC118 ||ROSINA_COPS_TOTAL_EMISSION_I_MT ||Total Emission I Microtip ||Total current through the Microtips ||Measured ||Std. ||Measurement with Ram Gauge ||2s ||1min ||< 20mA (TBC) ||3 ||5 || || |- |NRNAC119 ||ROSINA_COPS_P12V_I ||+12V I ||Current consumed by the +12V (includes current for +5V) ||Measured ||Std. ||Measurement ||2s ||1min ||< 180mA (TBC) ||3 ||5 || || |- |NRNAC11A || ||Spare || || || || || || || || || || || |- |NRNAC11B || ||COPS Ext. HK Header || || || || || || || || || || || |- | || ||Spare || || || || || || || || || || || |- |NRNAC11C ||ROSINA_COPS_ELECTR_TEMP ||Electr. Temp ||Temperature of the electronic (board2) ||Measured ||Ext. ||Measurement ||2s ||1min ||-30 to +60 DegC (TBC) ||3 ||TBD || || |- |NRNAC11D ||ROSINA_COPS_SENSOR_TEMP ||Sensor Temp ||Temperature of the Ram Gauge ||Measured ||Ext. ||Measurement ||2s ||1min ||-30 to +60 DegC (TBC) ||3 ||TBD || || |- |NRNAC11E ||ROSINA_COPS_P12V_FILAMENT ||+12V Filament ||+12V power supply for Nude Gauge electronic ||Measured ||Ext. ||Measurement with Nude Gauge ||2s ||1min ||Nominal value +/- 3% ||3 ||5 || || |- |NRNAC11F ||ROSINA_COPS_M12V_FILAMENT ||-12V Filament ||-12V power supply for Nude Gauge electronic ||Measured ||Ext. ||Measurement with Nude Gauge ||2s ||1min ||Nominal value +/- 3% ||3 ||5 || || |- |NRNAC120 ||ROSINA_COPS_P12V_MICROTIP ||+12V Microtip ||+12V power supply for Ram Gauge electronic ||Measured ||Ext. ||Measurement with Ram Gauge ||2s ||1min ||Nominal value +/- 3% ||3 ||5 || || |- |NRNAC121 ||ROSINA_COPS_M12V_MICROTIP ||-12V Microtip ||-12V power supply for Ram Gauge electronic ||Measured ||Ext. ||Measurement with Ram Gauge ||2s ||1min ||Nominal value +/- 3% ||3 ||5 || || |}