DATE ,TIME,INTLs,KEYWORD,PURPOSE,ACTIVITY DETAILS 20/6/01,17:30:00,,EU Shake,CTS Failure,Start to remove the EU from plate and transport to 306-221B for trobleshooting. ,,,,, 22/6/01,8:30:00,PK/BI,,,SBEU and USO unit removed from copper plate and transported to mag mapping. Sensor unit removed. Friday,,,,,Configuration: EU and CTS mounted on copper plate. J4 pin 3 shorted to chassis. EU J 9 test board installed with test cable to CTS J 12. Cable for CTS hskp in not connected ,,,,, 22/6/01,9:06:00,,Power on 302,,28v current .46A Friday,,,,, ,9:14:55,PK,,,cmd CTS dual continuum ,,,,,Testing CTS. ,,,,, ,9:19:00,,Power off 302,, ,,,,,Remove test board connector screws so that connector can mate better to EU J9. ,9:27:30,,Power on 303,,28V current = .46A ,9:31:51,,,,cmd CTS dual continuum. ,9:35:00,,Power off 303,, ,,,,,Demate tests cable from test board. ,9:48:00,,Power on 304,,28V current = .47A ,9:51:21,,,,cmd CTD dual continuum ,9:51:40,,,,data storage interupt ,9:52:50,,Power off 304,, ,,,,,Connect pin 2 on test board which was disconnected. ,10:15:00,,Power on 305,,28V current = .46A ,10:17:30,,,,cmd CTS dual continuum. ,10:24:07,,,,cmd ENG. ,,,,,Software change: set delay between clk to 25ms from 30sec. ,10:25:47,,,,cmd CTS dual continuum. ,10:30:34,,,,cmd ENG. ,10:32:00,,Power off 305,, ,10:33:00,,Power on 306,,Loading s/w to EEPROM: vxWorks_062201_clock5.bin ,,,,, ,,,,, ,,,,, ,10:49:00,,,,Code is running. ,10:50:15,,,,cmd CTS dual continuum. ,10:51:00,,,,3.3V_Curr = .6 amps. ,10:56:00,,Power off 306,,failure after CTS Run cmd issued ,10:57:00,,Power on 307,,Iin = .46 amps. ,11:01:17,,,,cmd CTS dual continuum. ,,,,,3.3 V I = .625 amps. Then .591 amps. ,11:05:15,,Power off 307,,failure after CTS Run cmd issued ,,,,,"remove test cable, install flight cable with test board" ,11:14:30,,Power on 308,, ,11:16:51,,,,cmd Cts Dual continuum. ,,,,,3.3 V I= .827 then .625 amps. ,11:19:13,,Power Off 308,, ,11:22:50,,Power on 309,, Removed Test board and connect Flt cable between CTS-J12 and EU-J9 . 28V current = .46A ,11:25:48,,mode change,,CTS-Dual ,11:33:55,,Power off 309,, ,11:34:30,,Power on 310,,28V current = .46A Loading S/W ,,,,,Loading eeprom with vxWorks_062001_clock3.bin to see if failure is the same as the one on 06-20 at 15:44. ,11:51:11,,mode change,,CTS/Dual ,12:16:43,,Power off 310,, ,12:17:00,,Power on 311,,Loading eeprom: vxWorks_062201_clock6.bin ,12:39:37,,,,cmd CTS dual continuum. ,12:49:00,,,,ENG ,12:49:21,,Power off 311,, ,,,,,Install test board between EU-J9 and flt cable P9 ,12:54:55,,Power on 312,, ,12:57:31,,,,cmd CTS dual continuum. ,13:10:10,,Power off 312,, ,13:12:30,,Power on 313,, ,13:18:35,,,,cmd CTS dual continuum. ,13:30:40,,Power off 313,, 23/6/01,,,,, Saturday,,,,, ,14:05:50,,Power on 314,,28.4V @ .46A ,14:09:59,,mode change,,CTS/Dual ,14:21:40,,Power off 314,, ,15:30:00,,,,Paul arrived. ,,,,,Data cable disconnected from EU elc. Pauls equip. connected to CTS. ,16:10:00,,,,Now operating from Chris's procedures. ,16:14:00,,Power on 315,,Iin = .48 amps. ,16:18:36,,,,cmd CTS dual continuum. ,,,,,Iin = .88 amps. ,,,,, ,18:15:00,,Power off 315,,Results of testing showed that instrument performed the same with Paul's lab top computer commanding the CTS. ,,,,, ,,,,,Reconnected data cable to EU ,,,,, 25/6/01,,,,, Monday,16:10:00,,,,Trying more CTS tests - will add code to try turning on 3 data bits at a time and see if failure occurs ,,,,, ,16:43:53,,Power on 316,,Loading eeprom with vxWorks_062501_clock7.bin ,16:59:53,,cmd,,CTS/Dual Continuum ,,,,,"Sent the CTS a clocks off command. Waited 3 seconds. Sent it a clocks on command. Waited 3 more seconds. Sent 3 commands separated by 1/10 second each and all 3 worked, unexpectedly! This has always failed in the past with 30 second waits between commands. Then, after 5 minutes we tried to send another command and got a DSI on the first command we tried to send it." ,,,,,"When looked at 3.3 V data, current never went high when all bits set ON but all commands accepted (didn't cause DSI)" ,17:22:26,,Power off 316,, ,,,,, 26/6/01,,,,, Tuesday,10:10:09,,Power on 317,,Loading eeprom with vxWorks_062601_clock8.bin. This will allow for 2 variable delays between clocks on/off and also between commands. ,10:26:25,,cmd,,CTS/Dual Continuum ,,,,,"Sent the CTS a clocks off command. Waited 3 seconds. Sent it a clocks on command. Waited 3 more seconds. Sent 3 commands separated by 1/10 second each and all 3 worked. This has always failed in the past with 30 second waits between commands. Then, after 3 seconds we tried to send another command and got a DSI on the first command we tried to send it." ,10:42:37,,Power off 317,, ,,,Config.,,Complete instrument install on copper plate after Mag . ,,,,,mapping. ,,,,, ,15:33:00,,Power on 318,, ,,,,,Robert is loading new software into the eeprom: vxWorks_061401_ee.bin. Last working software prior to diagnostic additions. ,15:51:57,,,,Cmd to cts /dual cont. ,15:56:50,,,,smm gunn volt = 4 ,15:58:32,,,,smm gunn volt = 5 ,15:58:54,,,,smm gunn volt = 3 ,16:03:17,,mode change,,ENG ,16:04:20,,Power off 318,, ,,,,, ,16:20:50,,power on 319,, ,16:39:27,,,,CMD to cts/dual cont ,,,,,Running software to send multiple commands after setting all clocks off and all clocks on. Are able to send multiple commands (for about 1 minute) before the DSI happened. ,16:59:00,,power off 319,, ,17:59:58,,Power on 320,, ,17:02:16,,cmd,,CTS Dual Continuum ,,,,,Running software to send multiple commands after setting all clocks off and all clocks on. This time was able to send multiple commands for about 22 seconds before the DSI happened. ,17:09:15,,power off 320,, ,17:09:25,,power on 321,, ,17:10:33,,cmd,,CTS Dual Continuum ,,,,,Running software to send multiple commands after setting all clocks off and all clocks on. This time was able to send multiple commands for about 2 seconds before the DSI happened. ,17:11:44,,power off 321,, ,17:12:31,,power on 322,, ,17:16:31,,cmd,,CTS Dual Continuum ,17:19:41,,Power off 322,, ,17:20:22,,Power on 323,, ,17:21:55,,mode change,,CTS/Dual ,,,,,Running software to send multiple commands after setting all clocks off and all clocks on. This time was able to send multiple commands for about 23 seconds before the DSI happened. ,17:23:34,,Power off 323,, ,17:24:16,,Power on 324,, ,17:25:14,,mode change,,CTS/Dual ,,,,,Running software to send multiple commands after setting all clocks off and all clocks on. This time was able to send multiple commands for about 4 seconds before the DSI happened. ,17:26:29,,power off 324,, ,17:27:17,,power on 325,, ,17:30:18,,cmd,,CTS Dual Continuum ,,,,,Running software to send multiple commands after setting all clocks off and all clocks on. This time was able to send multiple commands for about 16 seconds before the DSI happened. ,17:40:00,RN,Power off 325,, ,,,,, ,10:45:20,,Power on 326,,Testing new SIS s/w version 2 ,11:08:30,,Power off 326,, ,11:14:29,,Power on 327,,Test s/w loaded in eeprom: vxWorks_062701_pm12.bin ,11:33:34,,cmd,,CTS Warmup low level ,11:34:51,,mode change,,CTS/DUAL ,11:35:00,,Power off 327,, ,11:36:00,,Power on 328,, ,11:38:40,,cmd,,CTS Warmup low level ,11:39:29,,mode change,,CTS/Dual ,,,,,Instrument died at time about turning off +/- 12V ,11:43:00,,Power off 328,, ,11:44:00,,Power on 329,, ,11:46:00,,cmd,,CTS Warm low level ,11:47:25,,mode change ,,CTS Dual - ,11:57:00,,Power off 329,,Instrument died at time ablout turning off +/- 12V ,12:39:00,,Power on 330,,Test s/w loaded in eeprom: vxWorks_062701_pm12_2.bin ,12:59:42,,cmd,,CTS Warmup low level ,13:00:54,,mode change ,,CTS/Dual ,13:02:00,,Power off 330,, ,13:04:00,,Power on 331,, ,13:06:27,,cmd,,CTS Warmup low level ,13:07:12,,mode change ,,CTS/Dual ,13:24:33,,mode change ,,ENG ,13:26:05,,cmd,,CTS Warmup low level ,13:27:22,,mode change ,,CTS/Dual ,13:33:09,,mode change ,,ENG ,13:34:20,,cmd,,CTS Warmup low level ,13:36:38,,mode change ,,CTS/Dual ,13:39:00,,Power off 331,, ,13:39:30,,Power on 332,, ,13:58:00,,Power off 332,, ,13:58:30,,Power on 333,,s/w loaded in eeprom vxWorks_062701_ee.bin for TC/TM packet verification test by Peter. Nominal code. ,14:17:00,,Power off 333,, ,14:24:00,,Power on 334,,Start of Peter's Test ,15:36:28,,cmd,,CTS internal calibration (in order to check rejection s/w) ,16:06:23,,cmd,,CTS Warmup level off ,16:09:55,,cmd,,CTS Warmup level off ,16:16:00,,Power off 334,, ,16:22:00,,Power on 335,, ,16:28:00,,Power off 335,, ,16:30:00,,Power on 336,, ,16:51:00,,Power off 336,, ,16:53:00,,Power on 337,,Loading RAM with vxWorks_062701_obfifo.bin to diagnose state of OBFIFO HW both before and after reset ,17:28:18,,Power off 337,, ,17:33:20,,Power on 338,, ,17:35:00,,Power off 338,, ,18:10:00,,Power on 339,, ,18:12:20,,Power off 339,, ,18:12:50,,Power on 340,, ,18:15:00,,Power off 340,, ,18:15:20,,Power on 341,, ,18:17:10,,Power off 341,, ,18:17:30,,Power on 342,, ,18:52:00,,Power off 342,, ,18:53:12,,Power on 343,, ,18:56:12,,Power off 343,, ,18:56:45,,Power on 344,, ,19:02:25,,Power off 344,, ,19:02:54,,Power on 345,, ,19:24:00,,Power off 345,, 28/6/01,10:33:50,RN,Power on 346,,No 0xFEFE observed at SIS from HW ,10:35:48,RN,Power off 346,, ,10:36:20,RN,Power on 347,,0xFEFE observed ,10:39:24,RN,Power off 347,, ,10:40:34,RN,Power on 348,, ,10:47:23,RN,Power off 348,, ,11:05:53,RN,Power on 349,,Advanced time synchronization test ,11:08:10,RN,cmd,,CTS/Dual Continuum ,12:27:55,RN,cmd,,"CTS/Dual Continuum, intercepted at SIS and not sent to instrument" ,13:56:14,RN,Power off 349,, ,14:01:15,RN,Power on 350,, ,14:34:30,RN,Power off 350,, ,14:37:17,RN,Power on 351,, ,15:10:00,RN,Power off 351,, ,15:33:25,RN,Power on 352,, ,17:47:00,,Power off 352,,demate primary power cable and connected redundant at SIS unit. ,,,,, ,17:49:25,,Power on 353,,28.4V @ .66A ,18:08:05,,Power off 353,, ,18:10:00,,,,demate redundant power cable and connected primary at SIS unit. 29/6/01,,,,, Friday,,,,,Dry run of bench test ,8:42:50,,Power on 354,,28.4V @ .66A ,8:49:05,,mode change,,mm cont. ,8:52:27,,mode change,,smm cont. ,8:56:02,,mode change,,dual cont. ,8:59:26,,mode change,,CTS/smm cont ,9:05:12,,mode change,,CTS/dual ,9:11:33,,cmd,,smm gunn volt auto disable ,9:12:15,,cmd,,smm gunn volt = 4 ,9:14:33,,cmd,,ifp pwr contl = 0 ,9:17:45,,cmd,,cal htr = on ,9:19:31,,cmd,,cal htr = off ,9:20:45,,cmd,,USO = off ,9:21:35,,cmd,,USO = on ,9:22:32,,cmd,,mir pos = hot ,9:23:46,,cmd,,mir pos = cold ,9:24:45,,cmd,,mir pos = space ,9:26:02,,cmd,,CTS inter cal ,9:29:37,,cmd,,cont. subtraction 500 mm & 500 smm ,9:31:20,,cmd,,cont. subtraction 0 mm & 0 smm ,9:45:32,,cmd,,mm lna off ,9:50:27,,cmd,,mm lna on ,9:51:10,,cmd,,smm lna off ,9:51:50,,cmd,,smm lna on ,9:52:41,,cmd,,pll reset off ,9:53:34,,cmd,,pll reset on ,9:54:58,,cmd,,pll reset off ,9:55:52,,cmd,,CTS htr contl 40 Deg C low pwr ,9:59:45,,cmd,,disable sci data ,10:00:24,,cmd,,disable hskp data ,10:01:02,,cmd,,enable hskp ,10:02:10,,cmd,,enable sci ,10:03:05,,cmd,,connection test request ,10:04:30,,cmd,,reset tele output buffer ,10:05:17,,cmd,,time update ,10:35:32,,cmd,,ENG MODE ,10:43:05,,Power off 354,, ,14:33:00,,,,Start to shutdown all suport equipement and getting ready to move to 306-221 B on Monday 7-2-01 ,,,,, ,,,,, 3/7/01,13:24:00,TCK,,,All support equipment down in 306-221. ,,,,,"SU, USO, EU-CTS.digital, SBEU mounted on IHF (instrument handling fixture) on Optical table in front of NFR for smm NFR measurement" ,,,,, ,,Pstek,,,Setting up external bias for smmRFE for smm NFR AIDS#216861; Note: only using SU ,,,,, ,,TCK,,,See NFR_log04.txt for NFR testing details. ,,,,, ,18:00:00,TCK,,,Completed 4 LSB and 4 USB smmNFR patterns. ,,,,, 5/7/01,,,,,"configuration CTS, EU (full up instrument)" ,,,,, ,16:37:45,,Power on 355,,28.4V .66A ,,,,,Receiving eng data. ,16:49:00,AP,cmd,,CTS/Dual mode ,,,,, ,,TCK,,,Note: smmRFE IF W13 is flex w/ 15 dB pad cnt=6926 ,,TCK,,,Note: mmRFE IF W12 is flex w/ 6 dB pad cnt=7195 ,,,,, ,16:52:33,AP,cmd,,"CTS heater: 40C, low" ,16:53:32,AP,cmd,,calibration heater ON ,,,,,Twload1=20.31C Twload2=20.27C ,,,,, ,18:18:00,TCK,,,program attn in mmRFE line no extra pad ,,,,,"set attn to 5 dB, mm cnts=7734" ,,,,, ,18:20:10,TCK,cmd,,CTS internal calib ,18:24:27,TCK,cmd,,CTS/Dual (instr calib) ,,,,, ,,,,,Note: overs are on 2nd and 1st mirrors! ,,,,, ,18:27:30,TCK,,,"mm attn = 0, mm cnts=15481" ,18:29:30,,,,"mm attn = 1, mm cnts=13534" ,18:31:40,,,,"mm attn = 2, mm cnts=12023" ,18:32:30,,,,"mm attn = 3, mm cnts=10352" ,18:33:45,,,,"mm attn = 4, mm cnts=9058" ,18:34:24,,,,"mm attn = 5, mm cnts=7733" ,18:35:25,,,,"mm attn = 6, mm cnts=6783" ,18:36:21,,,,"mm attn = 7, mm cnts=5795" ,18:37:20,,,,"mm attn = 8, mm cnts=5045" ,18:38:00,,,,"mm attn = 9, mm cnts=4357" ,18:39:00,,,,"mm attn = 10, mm cnts=3888" ,18:40:00,,,,mm attn = 5 ,18:40:29,,cmd,,mm LNA off ,,,,,mm cnts=1681 ,18:41:27,,cmd,,mm LNA on ,,,,, ,,,,,move programmable attn to smmRFE line plus 10 dB ,,,,,set mmRFE pad to 6 dB ,,,,, ,18:45:00,,,,smm attn 0 ,,,,, ,,,,,"smm attn = 0, smm cnts=14273" ,18:45:44,TCK,cmd,,CTS/Dual ,18:50:41,,,,"smm attn = 1, smm cnts=12081" ,18:51:00,,cmd,,CTS/Dual ,18:54:16,,,,"smm attn = 2, smm cnts=10306" ,18:54:39,,cmd,,CTS/Dual ,18:58:11,,,,"smm attn = 3, smm cnts=8677" ,18:58:58,,cmd,,CTS/Dual ,19:02:33,,,,"smm attn = 4, smm cnts=" ,19:02:50,,cmd,,CTS/Dual ,19:06:15,,,,"smm attn = 5, smm cnts=7396" ,19:06:24,,cmd,,CTS/Dual ,,,,,"smm attn = 6, smm cnts=5446" ,19:09:51,,cmd,,CTS/Dual ,,,,,"smm attn = 7, smm cnts=4696" ,19:13:22,,cmd,,CTS/Dual ,,,,,"smm attn = 8, smm cnts=4081" ,19:16:35,,cmd,,CTS/Dual ,,,,,"smm attn = 9, smm cnts=3610" ,19:19:29,,cmd,,CTS/Dual ,,,,,"smm attn = 10, smm cnts=3253" ,19:22:59,,cmd,,CTS/Dual ,,,,, ,19:26:36,,cmd,,smm LNA off ,,,,,smm cnts=1755 ,19:29:48,,cmd,,smm LNA on ,,,,, ,19:31:03,TCK,cmd,,engineering mode ,,,,, ,19:47:00,TCK,Power OFF 355,, ,,,,, ,,,,, ,,,,, 6/7/01,9:58:40,TCK/AP,Power ON 356,,looking normal... 6/7/01,,,,, 6/7/01,,TCK,,,continuing linearity testing of IFP channels with injected noise sources 6/7/01,,,,, 6/7/01,,TCK,,,Note: using same components as in T/V linearity testing 6/7/01,,,,, 6/7/01,,,config,,smmRFE 10+5 dB fixed pad 6/7/01,,,,, 6/7/01,,,config,,mmRFE=0..11 dB attn=coupler=IFP 6/7/01,,,,,coupler=0..70 dB attn=mm Noise Source 6/7/01,,,,,mm Noise Source is Avantek plus 10+1+1 dB 6/7/01,,,,, 6/7/01,10:00:50,TCK/AP,cmd,,CTS/dual mode 6/7/01,,,,, 6/7/01,,,,,"mm attn= 6, noise attn=70 dB, mm cnts=6675" 6/7/01,,,,,"mm attn= 6, noise attn= 0 dB, mm cnts=7177" 6/7/01,,,,, 6/7/01,10:09:15,,,,"mm attn= 4, noise attn=70 dB, mm cnts=8849" 6/7/01,10:11:29,,,,"mm attn= 4, noise attn= 0 dB, mm cnts=9276" 6/7/01,,,,, 6/7/01,10:12:55,,,,"mm attn= 2, noise attn=70 dB, mm cnts=11727" 6/7/01,10:14:38,,,,"mm attn= 2, noise attn= 0 dB, mm cnts=12077" 6/7/01,,,,, 6/7/01,10:16:16,,,,"mm attn= 0, noise attn=70 dB, mm cnts=15128" 6/7/01,10:17:30,,,,"mm attn= 0, noise attn= 0 dB, mm cnts=15388" 6/7/01,,,,, 6/7/01,10:18:32,,,,"mm attn= 8, noise attn=70 dB, mm cnts=4910" 6/7/01,10:19:37,,,,"mm attn= 8, noise attn= 0 dB, mm cnts=5481" 6/7/01,,,,, 6/7/01,10:20:34,,,,"mm attn=10, noise attn=70 dB, mm cnts=3793" 6/7/01,10:21:30,,,,"mm attn=10, noise attn= 0 dB, mm cnts=4399" 6/7/01,,,,, 6/7/01,10:22:50,,,,"mm attn= 6, noise attn=70 dB, mm cnts=6567" 6/7/01,10:24:44,,,,"mm attn= 6, noise attn= 0 dB, mm cnts=7083" 6/7/01,,,,, 6/7/01,10:26:26,,,,"mm attn= 6, noise attn=70 dB" 6/7/01,,,,, 6/7/01,10:27:23,AP,,,mm LNA off 6/7/01,,,,,mm cnts=1679 6/7/01,10:28:32,AP,,,mm LNA on 6/7/01,,,,, 6/7/01,,,config,,mmRFE with 6 dB fixed pad 6/7/01,,,,, 6/7/01,,,config,,smmRFE=0..11 dB attn=coupler=IFP 6/7/01,,,,,coupler=0..70 dB attn=smm Noise Source 6/7/01,10:39:10,,,,smm Noise Source is JCA amp plus 3 dB 6/7/01,,,,, 6/7/01,10:39:10,,,,"smm attn= 6, noise attn=70 dB, smm cnts=4917" 6/7/01,10:41:02,,,,"smm attn= 6, noise attn= 0 dB, smm cnts=6411" 6/7/01,,,,, 6/7/01,10:45:00,,config,,add 6 dB total 3+6 dB on smm Noise source 6/7/01,,,,, 6/7/01,,,,,"smm attn= 6, noise attn= 0 dB, smm cnts=5298" 6/7/01,10:46:25,,,,"smm attn= 6, noise attn=70 dB, smm cnts=4912" 6/7/01,10:51:48,,cmd,,CTS/Dual 6/7/01,10:54:35,,,,"smm attn= 6, noise attn= 0 dB, smm cnts=5292" 6/7/01,10:54:59,,cmd,,CTS/Dual 6/7/01,,,,, 6/7/01,10:58:20,,cmd,,"CTS heater 40C, high" 6/7/01,10:58:52,0:15:00,cmd,,calibration load heater ON 6/7/01,11:13:17,0:03:00,cmd,,CTS internal calib 6/7/01,,,,, 6/7/01,11:16:47,,,,"smm attn= 4, noise attn=70 dB, smm cnts=6583" 6/7/01,11:17:21,0:03:00,cmd,,CTS/Dual 6/7/01,11:20:20,,,,"smm attn= 4, noise attn= 0 dB, smm cnts=6958" 6/7/01,11:20:52,0:03:00,cmd,,CTS/Dual 6/7/01,,,,, 6/7/01,11:23:53,,,,"smm attn= 0, noise attn=70 dB, smm cnts=12713 space" 6/7/01,11:24:16,0:03:00,cmd,,CTS/Dual 6/7/01,11:27:23,,,,"smm attn= 0, noise attn= 0 dB, smm cnts=13035 cold" 6/7/01,11:27:39,0:03:00,cmd,,CTS/Dual 6/7/01,,,,, 6/7/01,11:30:50,,,,"smm attn=10, noise attn=70 dB, smm cnts=3024 cold" 6/7/01,11:31:14,0:03:00,cmd,,CTS/Dual 6/7/01,11:35:00,,,,"smm attn=10, noise attn= 0 dB, smm cnts=3434 cold?" 6/7/01,11:35:11,0:03:00,cmd,,CTS/Dual 6/7/01,,,,, 6/7/01,11:38:01,,,,"smm attn= 6, noise attn=70 dB, smm cnts=4889" 6/7/01,11:38:18,0:03:00,cmd,,CTS/Dual 6/7/01,11:41:05,,,,"smm attn= 6, noise attn= 0 dB, smm cnts=5277" 6/7/01,11:41:23,0:03:00,cmd,,CTS/Dual 6/7/01,,,,, 6/7/01,11:44:22,,,,"smm attn= 6, noise attn=70 dB" 6/7/01,,,,, 6/7/01,11:44:59,0:01:00,cmd,,smm LNA off 6/7/01,11:45:16,0:03:00,cmd,,CTS/Dual 6/7/01,,,,,smm cnts=1755 6/7/01,11:50:08,0:01:00,cmd,,smm LNA on 6/7/01,,,,,empty telemetry queue... 6/7/01,12:05:54,AP/TCK,cmd,,engineering mode 6/7/01,,,,, 6/7/01,12:10:05,AP,Power OFF 356,,Finish with TK test. 6/7/01,,,,, 6/7/01,,,config,, 6/7/01,13:07:45,,Power on 357,,RN is running new code. EEPROM loaded with vxWorks_070601_ee.bin to test engineering HK packet change. ,13:44:06,RN,Power off 357,, ,13:46:00,RN,Power on 358,,Loading new software into EEPROM. vxWorks_070601_ee_2.bin